Determination on baicalin in Xiaoer Tuire oral liquid by HPLC
Yeqing Chen
Abstract
Yeqing Chen
Abstract
Objective:to establish HPLC method for the determination of baicalin in the Xiaoer Tuire oral liquid. Method: the HPLC method was carried on C 18 column using method-water-codeine(45∶55∶0.2) as mobile phase. Detection wavelengthwas set at 315nm. Results: in the HPLC method the calibration curve for baicalin was linear over the range of 0.15~1.2 μg/mL. The average recovery was 98.07% with RSD=1.05%. Conclusions: the method is simple, convenient and reliablefor quality control of the liquids.
A significance statement is not available in the OpenAlex record.
A contribution statement is not available in the OpenAlex record.
Method details are not available in the OpenAlex metadata.
Findings are not separately available in the OpenAlex metadata.
Limitations are not available in the OpenAlex metadata.
Application details are not available in the OpenAlex metadata.
Objective:to establish HPLC method for the determination of baicalin in the Xiaoer Tuire oral liquid. Method: the HPLC method was carried on C 18 column using method-water-codeine(45∶55∶0.2) as mobile phase. Detection wavelengthwas set at 315nm. Results: in the HPLC method the calibration curve for baicalin was linear over the range of 0.15~1.2 μg/mL. The average recovery was 98.07% with RSD=1.05%. Conclusions: the method is simple, convenient and reliablefor quality control of the liquids.