Comparison of intraocular pressures at different parts of eyeballs by Tono-Pen applanation tonometer
Hao Yanyan
Abstract
Hao Yanyan
Abstract
Aim: To explore the difference in intraocular pressure(IOP) on different part of the eye ball measured by Tono-Pen applanation tonometer. Methods: IOP on central cornea,limbus and sclera of 61 cases(99 eyes) were measured by Tono-Pen applanation tonometer. Results: IOP on central cornea,limbus and sclera were (2.52±0.74) kPa, ( 2.63± 0.79) kPa, (4.46±1.49) kPa,respectively. There was no statistical difference between the IOP on central cornea and the limbus(P0.05). IOP on sclera was higher than that on corneal center,and they were positively correlated ( r= 0.542, P=0.000). The linear equation was Y=1.824+0.104X(X≤2.793 kPa) and Y=1.493+0.359(X 2.793 kPa) (Y:IOP on centralcornea;X:IOP on sclera).Conclusion: IOP on sclera measured by Tono-Pen is reliable;IOP measured by Tono-Pen at any part of cornea is accurate.
A significance statement is not available in the OpenAlex record.
A contribution statement is not available in the OpenAlex record.
Method details are not available in the OpenAlex metadata.
Findings are not separately available in the OpenAlex metadata.
Limitations are not available in the OpenAlex metadata.
Application details are not available in the OpenAlex metadata.
Aim: To explore the difference in intraocular pressure(IOP) on different part of the eye ball measured by Tono-Pen applanation tonometer. Methods: IOP on central cornea,limbus and sclera of 61 cases(99 eyes) were measured by Tono-Pen applanation tonometer. Results: IOP on central cornea,limbus and sclera were (2.52±0.74) kPa, ( 2.63± 0.79) kPa, (4.46±1.49) kPa,respectively. There was no statistical difference between the IOP on central cornea and the limbus(P0.05). IOP on sclera was higher than that on corneal center,and they were positively correlated ( r= 0.542, P=0.000). The linear equation was Y=1.824+0.104X(X≤2.793 kPa) and Y=1.493+0.359(X 2.793 kPa) (Y:IOP on centralcornea;X:IOP on sclera).Conclusion: IOP on sclera measured by Tono-Pen is reliable;IOP measured by Tono-Pen at any part of cornea is accurate.
Key concepts: Sclera, Medicine, Goldmann Applanation Tonometer, Ophthalmology, Cornea, Intraocular pressure, Significant difference, Applanation tonometry