Comparison of intraocular pressures at different parts of eyeballs by Tono-Pen applanation tonometer

Hao Yanyan

Open publisher page 0 citations

Abstract

Aim: To explore the difference in intraocular pressure(IOP) on different part of the eye ball measured by Tono-Pen applanation tonometer. Methods: IOP on central cornea,limbus and sclera of 61 cases(99 eyes) were measured by Tono-Pen applanation tonometer. Results: IOP on central cornea,limbus and sclera were (2.52±0.74) kPa, ( 2.63± 0.79) kPa, (4.46±1.49) kPa,respectively. There was no statistical difference between the IOP on central cornea and the limbus(P0.05). IOP on sclera was higher than that on corneal center,and they were positively correlated ( r= 0.542, P=0.000). The linear equation was Y=1.824+0.104X(X≤2.793 kPa) and Y=1.493+0.359(X 2.793 kPa) (Y:IOP on centralcornea;X:IOP on sclera).Conclusion: IOP on sclera measured by Tono-Pen is reliable;IOP measured by Tono-Pen at any part of cornea is accurate.

About this research paper

What this paper is about

Aim: To explore the difference in intraocular pressure(IOP) on different part of the eye ball measured by Tono-Pen applanation tonometer. Methods: IOP on central cornea,limbus and sclera of 61 cases(99 eyes) were measured by Tono-Pen applanation tonometer. Results: IOP on central cornea,limbus and sclera were (2.52±0.74) kPa, ( 2.63± 0.79) kPa, (4.46±1.49) kPa,respectively. There was no statistical difference between the IOP on central cornea and the limbus(P0.05). IOP on sclera was higher than that on corneal center,and they were positively correlated ( r= 0.542, P=0.000). The linear equation was Y=1.824+0.104X(X≤2.793 kPa) and Y=1.493+0.359(X 2.793 kPa) (Y:IOP on centralcornea;X:IOP on sclera).Conclusion: IOP on sclera measured by Tono-Pen is reliable;IOP measured by Tono-Pen at any part of cornea is accurate.

Why it matters

A significance statement is not available in the OpenAlex record.

Key contribution

A contribution statement is not available in the OpenAlex record.

Method / approach

Method details are not available in the OpenAlex metadata.

Main findings

Findings are not separately available in the OpenAlex metadata.

Limitations

Limitations are not available in the OpenAlex metadata.

Applications

Application details are not available in the OpenAlex metadata.

Available abstract

Aim: To explore the difference in intraocular pressure(IOP) on different part of the eye ball measured by Tono-Pen applanation tonometer. Methods: IOP on central cornea,limbus and sclera of 61 cases(99 eyes) were measured by Tono-Pen applanation tonometer. Results: IOP on central cornea,limbus and sclera were (2.52±0.74) kPa, ( 2.63± 0.79) kPa, (4.46±1.49) kPa,respectively. There was no statistical difference between the IOP on central cornea and the limbus(P0.05). IOP on sclera was higher than that on corneal center,and they were positively correlated ( r= 0.542, P=0.000). The linear equation was Y=1.824+0.104X(X≤2.793 kPa) and Y=1.493+0.359(X 2.793 kPa) (Y:IOP on centralcornea;X:IOP on sclera).Conclusion: IOP on sclera measured by Tono-Pen is reliable;IOP measured by Tono-Pen at any part of cornea is accurate.

Key concepts: Sclera, Medicine, Goldmann Applanation Tonometer, Ophthalmology, Cornea, Intraocular pressure, Significant difference, Applanation tonometry