2010Chinese Journal of Health Laboratory TechnologyRequires access

Simultaneous determination of melamine and cyanuric acid in the urine by LC-APCI-MS/MS

Ji Shen

Open publisher page 0 citations

Abstract

Objective:To establish a liquid chromatography-atmospheric pressure chemical ionization tandem mass spectrometric method(LC-APCI-MS/MS) for the determination of melamine and cyanuric acid in urine samples.Methods: Urine samples were mixed with isotopic labeled internal standards,15N3-melamine and 13C3-cyanuric acid,and diluted with acetonitrile.After centrifugation and separated on a Waters XBridgeTM Amide column,the analytes were detected by APCI-MS/MS with the multiple reaction monitoring(MRM) mode,positive for melamine and negative for cyanuric acid.Results: Calibration curves were linear in the range of 20 ng/ml~1000 ng/ml for melamine and cyanuric acid,both with the correlation coefficient better than 0.995.The intra and inter-batch precision(RSD) both less than 10%,and the accuracies and recoveries were both in the range of 89.9%~111.2%.Conclusion: The method is simple,sensitive and specific for the quantitative determination of melamine and cyanuric acid in the urine.

About this research paper

What this paper is about

Objective:To establish a liquid chromatography-atmospheric pressure chemical ionization tandem mass spectrometric method(LC-APCI-MS/MS) for the determination of melamine and cyanuric acid in urine samples.Methods: Urine samples were mixed with isotopic labeled internal standards,15N3-melamine and 13C3-cyanuric acid,and diluted with acetonitrile.After centrifugation and separated on a Waters XBridgeTM Amide column,the analytes were detected by APCI-MS/MS with the multiple reaction monitoring(MRM) mode,positive for melamine and negative for cyanuric acid.Results: Calibration curves were linear in the range of 20 ng/ml~1000 ng/ml for melamine and cyanuric acid,both with the correlation coefficient better than 0.995.The intra and inter-batch precision(RSD) both less than 10%,and the accuracies and recoveries were both in the range of 89.9%~111.2%.Conclusion: The method is simple,sensitive and specific for the quantitative determination of melamine and cyanuric acid in the urine.

Why it matters

A significance statement is not available in the OpenAlex record.

Key contribution

A contribution statement is not available in the OpenAlex record.

Method / approach

Method details are not available in the OpenAlex metadata.

Main findings

Findings are not separately available in the OpenAlex metadata.

Limitations

Limitations are not available in the OpenAlex metadata.

Applications

Application details are not available in the OpenAlex metadata.

Available abstract

Objective:To establish a liquid chromatography-atmospheric pressure chemical ionization tandem mass spectrometric method(LC-APCI-MS/MS) for the determination of melamine and cyanuric acid in urine samples.Methods: Urine samples were mixed with isotopic labeled internal standards,15N3-melamine and 13C3-cyanuric acid,and diluted with acetonitrile.After centrifugation and separated on a Waters XBridgeTM Amide column,the analytes were detected by APCI-MS/MS with the multiple reaction monitoring(MRM) mode,positive for melamine and negative for cyanuric acid.Results: Calibration curves were linear in the range of 20 ng/ml~1000 ng/ml for melamine and cyanuric acid,both with the correlation coefficient better than 0.995.The intra and inter-batch precision(RSD) both less than 10%,and the accuracies and recoveries were both in the range of 89.9%~111.2%.Conclusion: The method is simple,sensitive and specific for the quantitative determination of melamine and cyanuric acid in the urine.

Key concepts: Cyanuric acid, Melamine, Chromatography, Chemistry, Atmospheric-pressure chemical ionization, Detection limit, Urine, Liquid chromatography–mass spectrometry

Related papers

Back to paper searchBrowse research topicsOriginal source
Simultaneous determination of melamine and cyanuric acid in the urine by LC-APCI-MS/MS — Research Paper | ScholarLens