Establishment of self-detection system and its value of domestic-made cholinesterase kit

Enzhi Wang

Open publisher page 0 citations

Abstract

Objective To analyse the comparable and clinical acceptability of results of testing cholinesterase(CHE)in patients serum between detection system of Jilin Huiqiang CHE kit in the Roche Modular and Beckman-Coulter LX20 automatic biochemical analyzer and the Roche Modular of traceable reference.Method Basing on paper EP9-A of the Committee for Clinical Laboratory Standards(NCCLS),CHE concentration in patients fresh serum through 2 self-detection systems and Roche Modular reference detection system were tested.Linear regression and correlation analysis were done,and U-test to determine whether the data from two detection systems had a significant difference.Results As results of testing CHE in serum had better comparability between 2 self-detection systems and Roche Modular reference detection system.The correlation coefficient(γ) 0.99,there were no significant differences(P0.05).In addition,comparability in testing results between 2 self-detection systems was superior to Roche Modular reference detection system.Conclusion As results of testing CHE in serum between 2 self-detection systems and Roche Modular reference detection system is accepted.

About this research paper

What this paper is about

Objective To analyse the comparable and clinical acceptability of results of testing cholinesterase(CHE)in patients serum between detection system of Jilin Huiqiang CHE kit in the Roche Modular and Beckman-Coulter LX20 automatic biochemical analyzer and the Roche Modular of traceable reference.Method Basing on paper EP9-A of the Committee for Clinical Laboratory Standards(NCCLS),CHE concentration in patients fresh serum through 2 self-detection systems and Roche Modular reference detection system were tested.Linear regression and correlation analysis were done,and U-test to determine whether the data from two detection systems had a significant difference.Results As results of testing CHE in serum had better comparability between 2 self-detection systems and Roche Modular reference detection system.The correlation coefficient(γ) 0.99,there were no significant differences(P0.05).In addition,comparability in testing results between 2 self-detection systems was superior to Roche Modular reference detection system.Conclusion As results of testing CHE in serum between 2 self-detection systems and Roche Modular reference detection system is accepted.

Why it matters

A significance statement is not available in the OpenAlex record.

Key contribution

A contribution statement is not available in the OpenAlex record.

Method / approach

Method details are not available in the OpenAlex metadata.

Main findings

Findings are not separately available in the OpenAlex metadata.

Limitations

Limitations are not available in the OpenAlex metadata.

Applications

Application details are not available in the OpenAlex metadata.

Available abstract

Objective To analyse the comparable and clinical acceptability of results of testing cholinesterase(CHE)in patients serum between detection system of Jilin Huiqiang CHE kit in the Roche Modular and Beckman-Coulter LX20 automatic biochemical analyzer and the Roche Modular of traceable reference.Method Basing on paper EP9-A of the Committee for Clinical Laboratory Standards(NCCLS),CHE concentration in patients fresh serum through 2 self-detection systems and Roche Modular reference detection system were tested.Linear regression and correlation analysis were done,and U-test to determine whether the data from two detection systems had a significant difference.Results As results of testing CHE in serum had better comparability between 2 self-detection systems and Roche Modular reference detection system.The correlation coefficient(γ) 0.99,there were no significant differences(P0.05).In addition,comparability in testing results between 2 self-detection systems was superior to Roche Modular reference detection system.Conclusion As results of testing CHE in serum between 2 self-detection systems and Roche Modular reference detection system is accepted.

Key concepts: Medicine, Modular design, Spectrum analyzer, Correlation coefficient, Comparability, Statistics, Computer science, Mathematics

Related papers

Back to paper searchBrowse research topicsOriginal source
Establishment of self-detection system and its value of domestic-made cholinesterase kit — Research Paper | ScholarLens