Comparison and evaluation of the self-established assay system for adenosine deaminase
Sun Wang-wei
Abstract
Sun Wang-wei
Abstract
Objective To evaluate the comparability and clinical acceptability of the results of serum adenosine deaminase(ADA) measured by self-established assay system with Beijing Leadman ADA kits in Roche Modular DPP clinical chemistry analyzers compared with traceable Roche Modular P reference assay system.Methods The concentrations of ADA in 40 fresh outpatient serum samples were analyzed in duplicate with two self-established systems and traceable Roche Modular reference system.The linear regression analysis was calculated according to National Committee for Clinical Laboratory Standards CLSI document EP9-A files.Results The comparability of serum ADA measurement results by self-established systems compared with Roche Modular reference system was acceptable.The regression equation was Y=0.998 7X+0.090 2,and the correlation coefficient(r2) was 0.998 4.Conclusion The results of serum ADA measured by our self-established systems have good comparability compared with Roche Modular P reference system,and it can replace the reference system to detect clinical specimens,which could be acceptable by clinical application.
A significance statement is not available in the OpenAlex record.
A contribution statement is not available in the OpenAlex record.
Method details are not available in the OpenAlex metadata.
Findings are not separately available in the OpenAlex metadata.
Limitations are not available in the OpenAlex metadata.
Application details are not available in the OpenAlex metadata.
Objective To evaluate the comparability and clinical acceptability of the results of serum adenosine deaminase(ADA) measured by self-established assay system with Beijing Leadman ADA kits in Roche Modular DPP clinical chemistry analyzers compared with traceable Roche Modular P reference assay system.Methods The concentrations of ADA in 40 fresh outpatient serum samples were analyzed in duplicate with two self-established systems and traceable Roche Modular reference system.The linear regression analysis was calculated according to National Committee for Clinical Laboratory Standards CLSI document EP9-A files.Results The comparability of serum ADA measurement results by self-established systems compared with Roche Modular reference system was acceptable.The regression equation was Y=0.998 7X+0.090 2,and the correlation coefficient(r2) was 0.998 4.Conclusion The results of serum ADA measured by our self-established systems have good comparability compared with Roche Modular P reference system,and it can replace the reference system to detect clinical specimens,which could be acceptable by clinical application.
Key concepts: Adenosine deaminase, Comparability, External quality assessment, Modular design, Reference values, Computer science, Mathematics, Medicine