Comparison of routine test kits of creatinine in human serum with reversed-phase high-pressure liquid chromatographic assay

Xu Jing

Open publisher page 0 citations

Abstract

Objective Precision and accuracy of Creatinine determination by Reversed-Phase High-Pressure Liquid Chromatographic Assay (HPLC) was investigated.This method was used as the criterion to evaluate the accuracy of routine test kits.Methods Beckman LX20 measure system used kinetic Jaffe method.Vitros250 Chemistry System used Vitro dry chemistry test kits,and enzymatic method was on the Hitachi 7170 Automated Analyzer.As to HPLC,C18 was stationary phase,and 5-Fluorouracil (5-Fu) was internal standard.Creatinine was determined at 235 nm.Results The with-in-day CVs (n=5) and day-to-day CVs(n=20)of HPLC were 2.0% and 5.1%,respectively.The absolute recovery was 95.31%,and the relative recovery was 98.95%.The results of certified reference materials were consistent with the target value,and average deviation was -0.31%~1.35%.HPLC was served as the independent variable.When ereatinine was 200 μmol/L,Bias of Beckman LX20 system,Vitros dry chemistry system and enzymatic method were -19%~21%,-14%~21% and -30%~11%, respectively;when creatinine was 200 μmol/L,Bias of Beckman LX20 system,Vitros dry chemistry system and enzymatic method were -10%~13%,-13%~14% and -20%~10%,respectively.Bias of enzymatic method results was mostly negative,when creatinine was 70 μmol/L,some could be -30%. Conclusions The precision and accuracy of HPLC was perfect.Compared with HPLC,bias of Beckman LX20 and Vitros system were relatively little,but enzymatic method results were lower than HPLC,Beckman LX20 system and Vitros system,and when creatinine concentration 70 μmol/L,enzymatic method could not measure exactly.

About this research paper

What this paper is about

Objective Precision and accuracy of Creatinine determination by Reversed-Phase High-Pressure Liquid Chromatographic Assay (HPLC) was investigated.This method was used as the criterion to evaluate the accuracy of routine test kits.Methods Beckman LX20 measure system used kinetic Jaffe method.Vitros250 Chemistry System used Vitro dry chemistry test kits,and enzymatic method was on the Hitachi 7170 Automated Analyzer.As to HPLC,C18 was stationary phase,and 5-Fluorouracil (5-Fu) was internal standard.Creatinine was determined at 235 nm.Results The with-in-day CVs (n=5) and day-to-day CVs(n=20)of HPLC were 2.0% and 5.1%,respectively.The absolute recovery was 95.31%,and the relative recovery was 98.95%.The results of certified reference materials were consistent with the target value,and average deviation was -0.31%~1.35%.HPLC was served as the independent variable.When ereatinine was 200 μmol/L,Bias of Beckman LX20 system,Vitros dry chemistry system and enzymatic method were -19%~21%,-14%~21% and -30%~11%, respectively;when creatinine was 200 μmol/L,Bias of Beckman LX20 system,Vitros dry chemistry system and enzymatic method were -10%~13%,-13%~14% and -20%~10%,respectively.Bias of enzymatic method results was mostly negative,when creatinine was 70 μmol/L,some could be -30%. Conclusions The precision and accuracy of HPLC was perfect.Compared with HPLC,bias of Beckman LX20 and Vitros system were relatively little,but enzymatic method results were lower than HPLC,Beckman LX20 system and Vitros system,and when creatinine concentration 70 μmol/L,enzymatic method could not measure exactly.

Why it matters

A significance statement is not available in the OpenAlex record.

Key contribution

A contribution statement is not available in the OpenAlex record.

Method / approach

Method details are not available in the OpenAlex metadata.

Main findings

Findings are not separately available in the OpenAlex metadata.

Limitations

Limitations are not available in the OpenAlex metadata.

Applications

Application details are not available in the OpenAlex metadata.

Available abstract

Objective Precision and accuracy of Creatinine determination by Reversed-Phase High-Pressure Liquid Chromatographic Assay (HPLC) was investigated.This method was used as the criterion to evaluate the accuracy of routine test kits.Methods Beckman LX20 measure system used kinetic Jaffe method.Vitros250 Chemistry System used Vitro dry chemistry test kits,and enzymatic method was on the Hitachi 7170 Automated Analyzer.As to HPLC,C18 was stationary phase,and 5-Fluorouracil (5-Fu) was internal standard.Creatinine was determined at 235 nm.Results The with-in-day CVs (n=5) and day-to-day CVs(n=20)of HPLC were 2.0% and 5.1%,respectively.The absolute recovery was 95.31%,and the relative recovery was 98.95%.The results of certified reference materials were consistent with the target value,and average deviation was -0.31%~1.35%.HPLC was served as the independent variable.When ereatinine was 200 μmol/L,Bias of Beckman LX20 system,Vitros dry chemistry system and enzymatic method were -19%~21%,-14%~21% and -30%~11%, respectively;when creatinine was 200 μmol/L,Bias of Beckman LX20 system,Vitros dry chemistry system and enzymatic method were -10%~13%,-13%~14% and -20%~10%,respectively.Bias of enzymatic method results was mostly negative,when creatinine was 70 μmol/L,some could be -30%. Conclusions The precision and accuracy of HPLC was perfect.Compared with HPLC,bias of Beckman LX20 and Vitros system were relatively little,but enzymatic method results were lower than HPLC,Beckman LX20 system and Vitros system,and when creatinine concentration 70 μmol/L,enzymatic method could not measure exactly.

Key concepts: Chromatography, High-performance liquid chromatography, Chemistry, Creatinine, Relative standard deviation, Detection limit, Biochemistry

Related papers

Back to paper searchBrowse research topicsOriginal source
Comparison of routine test kits of creatinine in human serum with reversed-phase high-pressure liquid chromatographic assay — Research Paper | ScholarLens