2015Chinese Journal of Analysis LaboratoryRequires access

Determination of 16 rare earth elements in animal liver by inductively coupled plasma mass spectrometry with dynamic reaction cell

Meng Hua-y

Open publisher page 0 citations

Abstract

A novel method for the determination of 16 rare earth elements in animal livers by inductively coupled plasma mass spectrometry( ICP-MS) was developed. Samples were dealt with digestion under high pressure,with103 Rh as the internal standard element to overcome non-spectroscopic interference and ammonia was used as dynamic reaction gas to eliminate mass spectroscopic interference. Satisfactory linearities of standard curve were obtained with correlation coefficients over 0. 999. The detection limits were in the range of 0. 2 ~ 33. 3 ng / L. The method was tested using national standard substances and the results were in good agreement with the recommended values.

About this research paper

What this paper is about

A novel method for the determination of 16 rare earth elements in animal livers by inductively coupled plasma mass spectrometry( ICP-MS) was developed. Samples were dealt with digestion under high pressure,with103 Rh as the internal standard element to overcome non-spectroscopic interference and ammonia was used as dynamic reaction gas to eliminate mass spectroscopic interference. Satisfactory linearities of standard curve were obtained with correlation coefficients over 0. 999. The detection limits were in the range of 0. 2 ~ 33. 3 ng / L. The method was tested using national standard substances and the results were in good agreement with the recommended values.

Why it matters

A significance statement is not available in the OpenAlex record.

Key contribution

A contribution statement is not available in the OpenAlex record.

Method / approach

Method details are not available in the OpenAlex metadata.

Main findings

Findings are not separately available in the OpenAlex metadata.

Limitations

Limitations are not available in the OpenAlex metadata.

Applications

Application details are not available in the OpenAlex metadata.

Available abstract

A novel method for the determination of 16 rare earth elements in animal livers by inductively coupled plasma mass spectrometry( ICP-MS) was developed. Samples were dealt with digestion under high pressure,with103 Rh as the internal standard element to overcome non-spectroscopic interference and ammonia was used as dynamic reaction gas to eliminate mass spectroscopic interference. Satisfactory linearities of standard curve were obtained with correlation coefficients over 0. 999. The detection limits were in the range of 0. 2 ~ 33. 3 ng / L. The method was tested using national standard substances and the results were in good agreement with the recommended values.

Key concepts: Chemistry, Inductively coupled plasma mass spectrometry, Mass spectrometry, Analytical Chemistry (journal), Rare earth, Detection limit, Microwave digestion, Standard curve