Clinical evaluation of the RAMP200 fluorescence immunoassay analyzer

Wan Hai-ying

Open publisher page 0 citations

Abstract

Objective Abiding by the laboratory quality requirement,to evaluate the performance of RAMP200 fluorescence immunoassay analyzer.Methods According to Series of documents of the American Society of Clinical and Laboratory Standards Institute(CLSI),following six items:the total precision,functional sensitivity,analytical measurement range,reportable range,the deviation between the holes,biological reference interval of RAMP200 fluorescence immunoassay analyzer were verified and evaluated.Results When BNP concentration levels were 80.4 and 1 169.6 pg/mL,the total precisions were 4.9% and 4.02%.The functional sensitivity was 10.58 pg/mL.The analytical measurement range was from 10.2 pg/mL to 1 805.0 pg/mL.The clinical reportable range was from 10.58 pg/mL to 5 415.0 pg/mL.The maximum deviation between the hole was 14.27%.Biological reference interval was from 15 to 86 pg/mL(Vendor reference interval 100 pg/mL).Conclusion The parameters of RAMP200 fluorescence immunoassay analyzer are consistent with the parameters provided by the vendor,so it can fit the clinical quality demand and it is suitable for the POCT.

About this research paper

What this paper is about

Objective Abiding by the laboratory quality requirement,to evaluate the performance of RAMP200 fluorescence immunoassay analyzer.Methods According to Series of documents of the American Society of Clinical and Laboratory Standards Institute(CLSI),following six items:the total precision,functional sensitivity,analytical measurement range,reportable range,the deviation between the holes,biological reference interval of RAMP200 fluorescence immunoassay analyzer were verified and evaluated.Results When BNP concentration levels were 80.4 and 1 169.6 pg/mL,the total precisions were 4.9% and 4.02%.The functional sensitivity was 10.58 pg/mL.The analytical measurement range was from 10.2 pg/mL to 1 805.0 pg/mL.The clinical reportable range was from 10.58 pg/mL to 5 415.0 pg/mL.The maximum deviation between the hole was 14.27%.Biological reference interval was from 15 to 86 pg/mL(Vendor reference interval 100 pg/mL).Conclusion The parameters of RAMP200 fluorescence immunoassay analyzer are consistent with the parameters provided by the vendor,so it can fit the clinical quality demand and it is suitable for the POCT.

Why it matters

A significance statement is not available in the OpenAlex record.

Key contribution

A contribution statement is not available in the OpenAlex record.

Method / approach

Method details are not available in the OpenAlex metadata.

Main findings

Findings are not separately available in the OpenAlex metadata.

Limitations

Limitations are not available in the OpenAlex metadata.

Applications

Application details are not available in the OpenAlex metadata.

Available abstract

Objective Abiding by the laboratory quality requirement,to evaluate the performance of RAMP200 fluorescence immunoassay analyzer.Methods According to Series of documents of the American Society of Clinical and Laboratory Standards Institute(CLSI),following six items:the total precision,functional sensitivity,analytical measurement range,reportable range,the deviation between the holes,biological reference interval of RAMP200 fluorescence immunoassay analyzer were verified and evaluated.Results When BNP concentration levels were 80.4 and 1 169.6 pg/mL,the total precisions were 4.9% and 4.02%.The functional sensitivity was 10.58 pg/mL.The analytical measurement range was from 10.2 pg/mL to 1 805.0 pg/mL.The clinical reportable range was from 10.58 pg/mL to 5 415.0 pg/mL.The maximum deviation between the hole was 14.27%.Biological reference interval was from 15 to 86 pg/mL(Vendor reference interval 100 pg/mL).Conclusion The parameters of RAMP200 fluorescence immunoassay analyzer are consistent with the parameters provided by the vendor,so it can fit the clinical quality demand and it is suitable for the POCT.

Key concepts: Spectrum analyzer, Immunoassay, Chromatography, Reproducibility, Reference range, Relative standard deviation, Fluorescence polarization immunoassay, Analytical Chemistry (journal)

Related papers

Back to paper searchBrowse research topicsOriginal source
Clinical evaluation of the RAMP200 fluorescence immunoassay analyzer — Research Paper | ScholarLens