Comparison Study of Reticulocyte Analysis Results Determined by Different Instruments

Qingyan Zhang

Open publisher page 0 citations

Abstract

Objective To explore the correlation of reticulocyte(RET)detection results between different instruments. Methods The reference method(Sysmex XE-5000)for instrument calibration was carried out,then the results of reticulocyte ratio tested by Sysmex XE-5000 and Sysmex XN-20 was compared. Results There was no statistical significance between two instruments for detecting reticulocyte ratio(P 0.05). Moreover,no statistically significant differences were found when comparing these data with the manual method(P 0.05). Conclusion Internal instruments are comparable for RET after calibrating Sysmex XE-5000 using reference method.

About this research paper

What this paper is about

Objective To explore the correlation of reticulocyte(RET)detection results between different instruments. Methods The reference method(Sysmex XE-5000)for instrument calibration was carried out,then the results of reticulocyte ratio tested by Sysmex XE-5000 and Sysmex XN-20 was compared. Results There was no statistical significance between two instruments for detecting reticulocyte ratio(P 0.05). Moreover,no statistically significant differences were found when comparing these data with the manual method(P 0.05). Conclusion Internal instruments are comparable for RET after calibrating Sysmex XE-5000 using reference method.

Why it matters

A significance statement is not available in the OpenAlex record.

Key contribution

A contribution statement is not available in the OpenAlex record.

Method / approach

Method details are not available in the OpenAlex metadata.

Main findings

Findings are not separately available in the OpenAlex metadata.

Limitations

Limitations are not available in the OpenAlex metadata.

Applications

Application details are not available in the OpenAlex metadata.

Available abstract

Objective To explore the correlation of reticulocyte(RET)detection results between different instruments. Methods The reference method(Sysmex XE-5000)for instrument calibration was carried out,then the results of reticulocyte ratio tested by Sysmex XE-5000 and Sysmex XN-20 was compared. Results There was no statistical significance between two instruments for detecting reticulocyte ratio(P 0.05). Moreover,no statistically significant differences were found when comparing these data with the manual method(P 0.05). Conclusion Internal instruments are comparable for RET after calibrating Sysmex XE-5000 using reference method.

Key concepts: Reticulocyte, Statistical analysis, Reference values, Calibration, Nuclear medicine, Medicine, Chemistry, Mathematics