Two different methods in detecting HIV antibody:a comparative study on results
Jun Huang
Abstract
Jun Huang
Abstract
OBJECTIVE To compave the efficiency of enzyme-Linked immunosorbent assay(ELISA) and colloidal gold immunochromatographic(GICA) to detect the sensitivity of HIV antibodies.METHODS The lowest limit of detection was compared by using ELISA and GICA reagents from 6 reagent suppliers.RESULTS All 6 samples showed positive results by using ELISA even when samples were diluted to tens of thousands of times while large difference existed in detection sensitivity by using different reagents in GICA.None of the reagents used in GICA was able to detect positive samples with OD value lower than 0.5 in ELISA.Only one reagent was able to detect samples with OD value above 0.5 while all the other reagents were able to detect only when OD value was above 1.5-2.0.CONCLUSION The sensitivity of ELISA is significantly higher than that of GICA.It advised to be more cautious when using GICA to detect HIV antibodies.
A significance statement is not available in the OpenAlex record.
A contribution statement is not available in the OpenAlex record.
Method details are not available in the OpenAlex metadata.
Findings are not separately available in the OpenAlex metadata.
Limitations are not available in the OpenAlex metadata.
Application details are not available in the OpenAlex metadata.
OBJECTIVE To compave the efficiency of enzyme-Linked immunosorbent assay(ELISA) and colloidal gold immunochromatographic(GICA) to detect the sensitivity of HIV antibodies.METHODS The lowest limit of detection was compared by using ELISA and GICA reagents from 6 reagent suppliers.RESULTS All 6 samples showed positive results by using ELISA even when samples were diluted to tens of thousands of times while large difference existed in detection sensitivity by using different reagents in GICA.None of the reagents used in GICA was able to detect positive samples with OD value lower than 0.5 in ELISA.Only one reagent was able to detect samples with OD value above 0.5 while all the other reagents were able to detect only when OD value was above 1.5-2.0.CONCLUSION The sensitivity of ELISA is significantly higher than that of GICA.It advised to be more cautious when using GICA to detect HIV antibodies.
Key concepts: Reagent, Detection limit, Antibody, Human immunodeficiency virus (HIV), Chromatography, Gold standard (test), Chemistry, Medicine