Graphite Furnace Atomic Absorption Spectrometry Determination of Aluminium in Viscera with Microwave Digestion Technique
HE Jin-lin
Abstract
HE Jin-lin
Abstract
Determination of aluminium in viscera is conducted by adopting microwave digestion technique. It is determined with nitric acid in a high - pressure microwave oven and then analyzed by graphite furnace atomic absorption spectrometry. The detection limit is 0. 91ng/ml,The calibration lines are linear in the range of 0-1000 ng/ml,and the recoveries of these materials range from 95. 1% to 104. 8% . The method is simple,accurate and low on the detection limit.
A significance statement is not available in the OpenAlex record.
A contribution statement is not available in the OpenAlex record.
Method details are not available in the OpenAlex metadata.
Findings are not separately available in the OpenAlex metadata.
Limitations are not available in the OpenAlex metadata.
Application details are not available in the OpenAlex metadata.
Determination of aluminium in viscera is conducted by adopting microwave digestion technique. It is determined with nitric acid in a high - pressure microwave oven and then analyzed by graphite furnace atomic absorption spectrometry. The detection limit is 0. 91ng/ml,The calibration lines are linear in the range of 0-1000 ng/ml,and the recoveries of these materials range from 95. 1% to 104. 8% . The method is simple,accurate and low on the detection limit.
Key concepts: Microwave digestion, Graphite furnace atomic absorption, Detection limit, Nitric acid, Atomic absorption spectroscopy, Microwave oven, Graphite, Aluminium