A New Non-Linearity Compensation Method for Pressure Sensors
Lu Hui
Abstract
Lu Hui
Abstract
On the basis of discussion of errir distribution of non linearity output of pressure sensors, this paper presents a new non linearity compensation method of pressure sensors. Also, it introduces the taking value method for three basic parameters (A、ω、φ) in error modification, i.e. compensation factors. Through the compensation of realistic output data of two pressure sensors, the non linearity degrees are reduced from 2.01% and 2.06% prior to the compensation to 0.69% and 0.33% posterior to the compensation respectively.
A significance statement is not available in the OpenAlex record.
A contribution statement is not available in the OpenAlex record.
Method details are not available in the OpenAlex metadata.
Findings are not separately available in the OpenAlex metadata.
Limitations are not available in the OpenAlex metadata.
Application details are not available in the OpenAlex metadata.
On the basis of discussion of errir distribution of non linearity output of pressure sensors, this paper presents a new non linearity compensation method of pressure sensors. Also, it introduces the taking value method for three basic parameters (A、ω、φ) in error modification, i.e. compensation factors. Through the compensation of realistic output data of two pressure sensors, the non linearity degrees are reduced from 2.01% and 2.06% prior to the compensation to 0.69% and 0.33% posterior to the compensation respectively.
Key concepts: Linearity, Compensation (psychology), Pressure sensor, Control theory (sociology), Computer science, Compensation methods, Electronic engineering, Engineering