Method Comparison and Bias Estimation of Different Testing Systems
Wei Wan-dong
Abstract
Wei Wan-dong
Abstract
Objective Several testing systems were compared to study if the results were in the allowable range.Methods Using HITACHI 7080/Roche reagent/cfas calibrator as reference system(X),Olympus 5400/Olympus reagent/Olympus calibrator(Y1),Olympus 5400/Zhongsheng reagent/Olympus calibrator(Y2),Olympus 2700/Zhongsheng reagent/Beckman calibrator(Y3)were three testing system.Based on the NCCLS EP9-A document,more than 8 samples per day were collected and 12 biochemical targets were analyzed with 4 testing systems respectively for 5 days.Linearity equation and the relativity coefficient were accumulated.According to the error allowable of CLIA'88 about the interlaboratory quality assessment,the results were judged.Result The relativity between testing system 2 and reference system was the best.Except for TG,the relativity between testing systing system l and reference system was good.The relativity between testing system 3 and reference system was also right,except P and CHE.TG,LDH and GLU(low consistence)could not be accepted by testing system 1,Testing system 2 could not accept CHO.Testing system 3 could accept all the eight items.Conclusion After comparison,the results beyond the reference range were still comparable between testing systems and conference system after calibrating.
A significance statement is not available in the OpenAlex record.
A contribution statement is not available in the OpenAlex record.
Method details are not available in the OpenAlex metadata.
Findings are not separately available in the OpenAlex metadata.
Limitations are not available in the OpenAlex metadata.
Application details are not available in the OpenAlex metadata.
Objective Several testing systems were compared to study if the results were in the allowable range.Methods Using HITACHI 7080/Roche reagent/cfas calibrator as reference system(X),Olympus 5400/Olympus reagent/Olympus calibrator(Y1),Olympus 5400/Zhongsheng reagent/Olympus calibrator(Y2),Olympus 2700/Zhongsheng reagent/Beckman calibrator(Y3)were three testing system.Based on the NCCLS EP9-A document,more than 8 samples per day were collected and 12 biochemical targets were analyzed with 4 testing systems respectively for 5 days.Linearity equation and the relativity coefficient were accumulated.According to the error allowable of CLIA'88 about the interlaboratory quality assessment,the results were judged.Result The relativity between testing system 2 and reference system was the best.Except for TG,the relativity between testing systing system l and reference system was good.The relativity between testing system 3 and reference system was also right,except P and CHE.TG,LDH and GLU(low consistence)could not be accepted by testing system 1,Testing system 2 could not accept CHO.Testing system 3 could accept all the eight items.Conclusion After comparison,the results beyond the reference range were still comparable between testing systems and conference system after calibrating.
Key concepts: Reagent, Theory of relativity, Range (aeronautics), Mathematics, Statistics, Computer science, Chemistry, Physics