Clinical application analysis on the performance of Sysmex UF-1000i automated urinary sediment analyzer
Luo Jia
Abstract
Luo Jia
Abstract
Objective To investigate the clinical application values of Sysmex UF-1000i automated urinary sediment analyzer.Methods 2 000 urine samples were determined by UF-1000i automated urinary sediment analyzer and microscopy.The results were contrasted,and the reasons of false positive in UF-1000i analyzer were analyzed.Results When using Sysmex UF-1000i analyzer,the detecting sensitivities of white blood cell(WBC),red blood cell(RBC),cast and small round epithelial cell were 99.2%,99.3%,94.7% and 95.1%,and the false positive rates of them were 5.8%,6.1%,10.4% and 4.6%,respectively.Many factors,like viscose rayon,crystals,bacteria,yeast-like fungus,small round epithelial cells,fat droplets and lecithin body may affect the determination of urine formed elements.Conclusions It is an effectual method to detect urinary sediments by Sysmex UF-1000i analyzer.It is very important to observe sediments by microscopy when all above interferences are positive or increase.
A significance statement is not available in the OpenAlex record.
A contribution statement is not available in the OpenAlex record.
Method details are not available in the OpenAlex metadata.
Findings are not separately available in the OpenAlex metadata.
Limitations are not available in the OpenAlex metadata.
Application details are not available in the OpenAlex metadata.
Objective To investigate the clinical application values of Sysmex UF-1000i automated urinary sediment analyzer.Methods 2 000 urine samples were determined by UF-1000i automated urinary sediment analyzer and microscopy.The results were contrasted,and the reasons of false positive in UF-1000i analyzer were analyzed.Results When using Sysmex UF-1000i analyzer,the detecting sensitivities of white blood cell(WBC),red blood cell(RBC),cast and small round epithelial cell were 99.2%,99.3%,94.7% and 95.1%,and the false positive rates of them were 5.8%,6.1%,10.4% and 4.6%,respectively.Many factors,like viscose rayon,crystals,bacteria,yeast-like fungus,small round epithelial cells,fat droplets and lecithin body may affect the determination of urine formed elements.Conclusions It is an effectual method to detect urinary sediments by Sysmex UF-1000i analyzer.It is very important to observe sediments by microscopy when all above interferences are positive or increase.
Key concepts: Urinary sediment, Spectrum analyzer, Urine, Urinary system, Chromatography, White blood cell, Pathology, Chemistry