[A comparison of the performances between the ionization chamber dosimeters used in radiation therapy].
Guoqing Li, Mu-yan Lin
Abstract
Guoqing Li, Mu-yan Lin
Abstract
The structural principles and technical performances of different ionization chamber dosimeters used in radiotherapy have been briefly introduced in this paper. In the mean time, a comparison of technical performances of four different kinds of ionization chamber dosimeter used in hospitals in China has also been made. The data show their quality meet the required specification specified by IEC731 and JJG912-96, so they can be used as field survey dosimeters in hospitals. Some precautions in using these dosimeters are given here too.
A significance statement is not available in the OpenAlex record.
A contribution statement is not available in the OpenAlex record.
Method details are not available in the OpenAlex metadata.
Findings are not separately available in the OpenAlex metadata.
Limitations are not available in the OpenAlex metadata.
Application details are not available in the OpenAlex metadata.
The structural principles and technical performances of different ionization chamber dosimeters used in radiotherapy have been briefly introduced in this paper. In the mean time, a comparison of technical performances of four different kinds of ionization chamber dosimeter used in hospitals in China has also been made. The data show their quality meet the required specification specified by IEC731 and JJG912-96, so they can be used as field survey dosimeters in hospitals. Some precautions in using these dosimeters are given here too.
Key concepts: Dosimeter, Ionization chamber, Ionization, Medical physics, Dosimetry, Quality (philosophy), Nuclear engineering, Materials science