Evaluation of the agilent 3458A time jitter performance
R. Lapuh, Boštjan Voljč, Matjaž Lindic
Abstract
R. Lapuh, Boštjan Voljč, Matjaž Lindic
Abstract
The paper describes the mechanism of Agilent 3458A sampling time jitter present when external triggering is used at synchronous two channel sampling. Based on measurements it is shown that the master DMM adds approximately 3.2 ns of its own time jitter while triggering the slave DMM and that the effective time jitter for externally triggered DMM remains at 7 ns, well below ±50 ns maximum jitter as specified by manufacturer.
OpenAlex reports 3 citations for this work. Citation counts describe recorded attention and do not establish research quality.
A contribution statement is not available in the OpenAlex record.
Method details are not available in the OpenAlex metadata.
Findings are not separately available in the OpenAlex metadata.
Limitations are not available in the OpenAlex metadata.
Application details are not available in the OpenAlex metadata.
The paper describes the mechanism of Agilent 3458A sampling time jitter present when external triggering is used at synchronous two channel sampling. Based on measurements it is shown that the master DMM adds approximately 3.2 ns of its own time jitter while triggering the slave DMM and that the effective time jitter for externally triggered DMM remains at 7 ns, well below ±50 ns maximum jitter as specified by manufacturer.
Key concepts: Jitter, Computer science, Sampling (signal processing), Channel (broadcasting), Sampling time, Electronic engineering, Engineering, Telecommunications