Comparison of Thresholds for Pulmonary Capillary Hemorrhage Induced by Pulsed-wave and B-mode Ultrasound
Douglas L. Miller, Chunyan Dou, Krishnan Raghavendran
Abstract
Douglas L. Miller, Chunyan Dou, Krishnan Raghavendran
Abstract
Pulsed ultrasound was found to induce pulmonary capillary hemorrhage (PCH) in mice about 25 years ago but remains a poorly understood risk factor for pulmonary diagnostic ultrasound. In early research using laboratory fixed beam ultrasound, thresholds for PCH had frequency variation from 1-4 MHz similar to the Mechanical Index. In recent research, thresholds for B mode diagnostic ultrasound from 1.5-12 MHz had little dependence on frequency. To compare the diagnostic ultrasound method to laboratory pulsed exposure, thresholds for fixed beam ultrasound were determined using comparable methods at 1.5 and 7.5 MHz. PCH thresholds were lower for simple fixed-beam pulse modes than for B mode and in approximate agreement with early research. However, for comparable timing parameters, PCH thresholds had little dependence on ultrasonic frequency. These findings suggest that the MI may not be directly useful as a dosimetric parameter for safety guidance in pulmonary ultrasound.
OpenAlex reports 2 citations for this work. Citation counts describe recorded attention and do not establish research quality.
A contribution statement is not available in the OpenAlex record.
Method details are not available in the OpenAlex metadata.
Findings are not separately available in the OpenAlex metadata.
Limitations are not available in the OpenAlex metadata.
Application details are not available in the OpenAlex metadata.
Pulsed ultrasound was found to induce pulmonary capillary hemorrhage (PCH) in mice about 25 years ago but remains a poorly understood risk factor for pulmonary diagnostic ultrasound. In early research using laboratory fixed beam ultrasound, thresholds for PCH had frequency variation from 1-4 MHz similar to the Mechanical Index. In recent research, thresholds for B mode diagnostic ultrasound from 1.5-12 MHz had little dependence on frequency. To compare the diagnostic ultrasound method to laboratory pulsed exposure, thresholds for fixed beam ultrasound were determined using comparable methods at 1.5 and 7.5 MHz. PCH thresholds were lower for simple fixed-beam pulse modes than for B mode and in approximate agreement with early research. However, for comparable timing parameters, PCH thresholds had little dependence on ultrasonic frequency. These findings suggest that the MI may not be directly useful as a dosimetric parameter for safety guidance in pulmonary ultrasound.
Key concepts: Mechanical index, Ultrasound, Diagnostic ultrasound, Ultrasonic sensor, Beam (structure), Pulse (music), Materials science, Pulse duration