2009Asian Journal of Research in ChemistryRequires access

Determination of Telmisartan in Solid Dosage Form by RP-HPLC

N. R. Vekariya, G. F. Patel, Rohit B Dholakiya

Open publisher page 4 citations

Abstract

A simple, precise, rapid and accurate reverse phase HPLC method developed for the estimation of Telmisartan in tablet dosage form. Luna 5 μ C18, 250 × 4.6 mm, particle size 5 μm, with mobile phase consisting of 5 mM Phosphate buffer: Acetonitrile (60:40, v/v), pH 7.4 was used. The flow rate was 1 ml/min and the effluents were monitored at 295 nm. The retention time was 7.02 min. The detector response was linear in the concentration of 2–14 μg/ml. The respective linear regression equation being Y=61480X-10188. The limit of detection and limit of quantification was 0.06 and 0.18 mcg/ml respectively. The percentage assay of Telmisartan was 100.28 ± 0.93%. The method was validated by determining its accuracy, precision and system suitability. The results of the study showed that the proposed RP-HPLC method is simple, rapid, precise and accurate, which is useful for the routine determination of Telmisartan in bulk drug and in its pharmaceutical dosage form.

About this research paper

What this paper is about

A simple, precise, rapid and accurate reverse phase HPLC method developed for the estimation of Telmisartan in tablet dosage form. Luna 5 μ C18, 250 × 4.6 mm, particle size 5 μm, with mobile phase consisting of 5 mM Phosphate buffer: Acetonitrile (60:40, v/v), pH 7.4 was used. The flow rate was 1 ml/min and the effluents were monitored at 295 nm. The retention time was 7.02 min. The detector response was linear in the concentration of 2–14 μg/ml. The respective linear regression equation being Y=61480X-10188. The limit of detection and limit of quantification was 0.06 and 0.18 mcg/ml respectively. The percentage assay of Telmisartan was 100.28 ± 0.93%. The method was validated by determining its accuracy, precision and system suitability. The results of the study showed that the proposed RP-HPLC method is simple, rapid, precise and accurate, which is useful for the routine determination of Telmisartan in bulk drug and in its pharmaceutical dosage form.

Why it matters

OpenAlex reports 4 citations for this work. Citation counts describe recorded attention and do not establish research quality.

Key contribution

A contribution statement is not available in the OpenAlex record.

Method / approach

Method details are not available in the OpenAlex metadata.

Main findings

Findings are not separately available in the OpenAlex metadata.

Limitations

Limitations are not available in the OpenAlex metadata.

Applications

Application details are not available in the OpenAlex metadata.

Available abstract

A simple, precise, rapid and accurate reverse phase HPLC method developed for the estimation of Telmisartan in tablet dosage form. Luna 5 μ C18, 250 × 4.6 mm, particle size 5 μm, with mobile phase consisting of 5 mM Phosphate buffer: Acetonitrile (60:40, v/v), pH 7.4 was used. The flow rate was 1 ml/min and the effluents were monitored at 295 nm. The retention time was 7.02 min. The detector response was linear in the concentration of 2–14 μg/ml. The respective linear regression equation being Y=61480X-10188. The limit of detection and limit of quantification was 0.06 and 0.18 mcg/ml respectively. The percentage assay of Telmisartan was 100.28 ± 0.93%. The method was validated by determining its accuracy, precision and system suitability. The results of the study showed that the proposed RP-HPLC method is simple, rapid, precise and accurate, which is useful for the routine determination of Telmisartan in bulk drug and in its pharmaceutical dosage form.

Key concepts: Telmisartan, Chromatography, Dosage form, Detection limit, Chemistry, High-performance liquid chromatography, Particle size, Acetonitrile

Related papers

Back to paper searchBrowse research topicsOriginal source
Determination of Telmisartan in Solid Dosage Form by RP-HPLC — Research Paper | ScholarLens