Agreement of airway resistance measurements by two different techniques of body-plethysmography and impulse oscillometry in asthmatic patients
Akbar Sharifi, Khalil Ansarin
Abstract
Akbar Sharifi, Khalil Ansarin
Abstract
Background: Airway resistance can be measured by different techniques of body-plethysmography and impulse oscillometry (IOS). So far there has been no systematic study comparing validity of these techniques in relation to clinical condition of the patients reported. Aims and objectives: We investigated correlation between these techniques in assessment of airway resistance and asthma control test. Methods: In 92 patients with asthma selected on the basis of ATS criteria for diagnosis of asthma and GINA asthma control test (ACT) questionnaire completed. Pulmonary function tests including body-plethysmography with airway resistance measurement and impulse oscillometry measuring total airway resistance at 5 Hz and 20Hz was done using IOS. Results: ACT score has a significant correlation with a r value of - 0.34 with total airway resistance measured by body-plethysmography (p=0.003) and also significant correlation with r value of - 0.31 (p=0.002) with airway resistance measured by IOS at 5Hz but no significant correlation with airway resistance measured by IOS at 20 Hz. There is a significant correlation between airway resistance measured by body plethysmography and airway resistance measured by IOS at 5 Hz and at 20 Hz. Conclusion: These finding indicates a good correlation of total airway resistance measured by body plethysmography and by IOS at 5 Hz as well as 20 Hz, which is stronger with the first, Also significant negative correlation between ACT score and airway resistance centered in peripheral airways measured by IOS at 5 Hz but not with the resistance at central airways measured mainly by IOS at 20Hz.
A significance statement is not available in the OpenAlex record.
A contribution statement is not available in the OpenAlex record.
Method details are not available in the OpenAlex metadata.
Findings are not separately available in the OpenAlex metadata.
Limitations are not available in the OpenAlex metadata.
Application details are not available in the OpenAlex metadata.
Background: Airway resistance can be measured by different techniques of body-plethysmography and impulse oscillometry (IOS). So far there has been no systematic study comparing validity of these techniques in relation to clinical condition of the patients reported. Aims and objectives: We investigated correlation between these techniques in assessment of airway resistance and asthma control test. Methods: In 92 patients with asthma selected on the basis of ATS criteria for diagnosis of asthma and GINA asthma control test (ACT) questionnaire completed. Pulmonary function tests including body-plethysmography with airway resistance measurement and impulse oscillometry measuring total airway resistance at 5 Hz and 20Hz was done using IOS. Results: ACT score has a significant correlation with a r value of - 0.34 with total airway resistance measured by body-plethysmography (p=0.003) and also significant correlation with r value of - 0.31 (p=0.002) with airway resistance measured by IOS at 5Hz but no significant correlation with airway resistance measured by IOS at 20 Hz. There is a significant correlation between airway resistance measured by body plethysmography and airway resistance measured by IOS at 5 Hz and at 20 Hz. Conclusion: These finding indicates a good correlation of total airway resistance measured by body plethysmography and by IOS at 5 Hz as well as 20 Hz, which is stronger with the first, Also significant negative correlation between ACT score and airway resistance centered in peripheral airways measured by IOS at 5 Hz but not with the resistance at central airways measured mainly by IOS at 20Hz.
Key concepts: Plethysmograph, Medicine, Airway resistance, Airway, Asthma, Correlation, Cardiology, Impulse (physics)