Simultaneous Estimation of Levofloxacin Hemihydrate and Ornidazole by High Performance Liquid Chromatography in Pharmaceutical Dosage Form

Manish Devgun, Anand Singh, Lovkesh Bhatia, S. M. Vijayendra Swamy, Sumedha Goyat

Open publisher page 0 citations

Abstract

A new simple, selective, rapid, precise and economical reverse phase isocratic High Performance Liquid Chromatography (HPLC-RP) method has been developed for the simultaneous estimation of Levofloxacin hemihydrate and Ornidazole in pharmaceutical preparations using Kromacil 5 μ 100 A C18, 150 x 4.6 mm column and the mobile phase consisted of acetonitrile:water (40:60), pH adjusted to 3.0 ± 0.05 with dil. Orthophosphoric acid. The flow rate was kept at 1.3 ml/min. The Ultraviolet Violet (U.V.) detection was done at 302 nm. The retention times for these drugs were 1.051 min and 1.777 min, respectively. The HPLC method was validated. The present method was found to be specific. The number of theoretical plates came out to be fairly high, i.e., 1082 and 2051 for the Levofloxacin hemihydrate and Ornidazole respectively. The tailing factor came out to be 1.26 and 1.11 for Levofloxacin hemihydrate and Ornidazole respectively. The relative standard deviation or precision was calculated to be 0.5 and 0.3 for Levofloxacin hemihydrate and Ornidazole respectively. The method was tested for its robustness as the resolution came out to be 5.09. With this present method, the calibration curve was found to be linear and run time was just 3 min., which made the method rapid and economical.

About this research paper

What this paper is about

A new simple, selective, rapid, precise and economical reverse phase isocratic High Performance Liquid Chromatography (HPLC-RP) method has been developed for the simultaneous estimation of Levofloxacin hemihydrate and Ornidazole in pharmaceutical preparations using Kromacil 5 μ 100 A C18, 150 x 4.6 mm column and the mobile phase consisted of acetonitrile:water (40:60), pH adjusted to 3.0 ± 0.05 with dil. Orthophosphoric acid. The flow rate was kept at 1.3 ml/min. The Ultraviolet Violet (U.V.) detection was done at 302 nm. The retention times for these drugs were 1.051 min and 1.777 min, respectively. The HPLC method was validated. The present method was found to be specific. The number of theoretical plates came out to be fairly high, i.e., 1082 and 2051 for the Levofloxacin hemihydrate and Ornidazole respectively. The tailing factor came out to be 1.26 and 1.11 for Levofloxacin hemihydrate and Ornidazole respectively. The relative standard deviation or precision was calculated to be 0.5 and 0.3 for Levofloxacin hemihydrate and Ornidazole respectively. The method was tested for its robustness as the resolution came out to be 5.09. With this present method, the calibration curve was found to be linear and run time was just 3 min., which made the method rapid and economical.

Why it matters

A significance statement is not available in the OpenAlex record.

Key contribution

A contribution statement is not available in the OpenAlex record.

Method / approach

Method details are not available in the OpenAlex metadata.

Main findings

Findings are not separately available in the OpenAlex metadata.

Limitations

Limitations are not available in the OpenAlex metadata.

Applications

Application details are not available in the OpenAlex metadata.

Available abstract

A new simple, selective, rapid, precise and economical reverse phase isocratic High Performance Liquid Chromatography (HPLC-RP) method has been developed for the simultaneous estimation of Levofloxacin hemihydrate and Ornidazole in pharmaceutical preparations using Kromacil 5 μ 100 A C18, 150 x 4.6 mm column and the mobile phase consisted of acetonitrile:water (40:60), pH adjusted to 3.0 ± 0.05 with dil. Orthophosphoric acid. The flow rate was kept at 1.3 ml/min. The Ultraviolet Violet (U.V.) detection was done at 302 nm. The retention times for these drugs were 1.051 min and 1.777 min, respectively. The HPLC method was validated. The present method was found to be specific. The number of theoretical plates came out to be fairly high, i.e., 1082 and 2051 for the Levofloxacin hemihydrate and Ornidazole respectively. The tailing factor came out to be 1.26 and 1.11 for Levofloxacin hemihydrate and Ornidazole respectively. The relative standard deviation or precision was calculated to be 0.5 and 0.3 for Levofloxacin hemihydrate and Ornidazole respectively. The method was tested for its robustness as the resolution came out to be 5.09. With this present method, the calibration curve was found to be linear and run time was just 3 min., which made the method rapid and economical.

Key concepts: Ornidazole, Chromatography, High-performance liquid chromatography, Chemistry, Levofloxacin, Dosage form, Resolution (logic), Retention time

Related papers

Back to paper searchBrowse research topicsOriginal source
Simultaneous Estimation of Levofloxacin Hemihydrate and Ornidazole by High Performance Liquid Chromatography in Pharmaceutical Dosage Form — Research Paper | ScholarLens