Doppler Angle and Flow Velocity Estimations Using the Classic and Transverse
Bor-Ray Lee, Huihua Kenny Chiang, Cheng‐Deng Kuo, Win-Li Lin, San-Kan Lee
Abstract
Bor-Ray Lee, Huihua Kenny Chiang, Cheng‐Deng Kuo, Win-Li Lin, San-Kan Lee
Abstract
Current clinical Doppler ultrasound systems could only measure the flow vector parallel to the ultra- sound beam axis, and the knowledge of the Doppler angle (beam-to-flow angle) is needed to calculate the real flow ve- locity. Currently, the Doppler angle is determined visually by manually aligning a vessel axis marker along the blood vessel on the duplex scan image of the ultrasound. The ap- plication of this procedure is often limited by practical con- straints; therefore, measurements are not reliable. In order to overcome this problem, we developed a simple Doppler angle and flow velocity estimation method using a combi- nation of the classic and transverse Doppler effects. This method uses only a single focused annular array transducer to estimate the Doppler angle and the flow velocity. We have verified experimentally that this method is successful for measuring constant flow in a flow phantom between 45O and 80' Doppler angle. The standard deviation of the esti- mated Doppler angles is less than 4.5O. This method could be implemented easily in medical Doppler ultrasound sys- tems to automatically estimate the Doppler angle and the flow velocity.
A significance statement is not available in the OpenAlex record.
A contribution statement is not available in the OpenAlex record.
Method details are not available in the OpenAlex metadata.
Findings are not separately available in the OpenAlex metadata.
Limitations are not available in the OpenAlex metadata.
Application details are not available in the OpenAlex metadata.
Current clinical Doppler ultrasound systems could only measure the flow vector parallel to the ultra- sound beam axis, and the knowledge of the Doppler angle (beam-to-flow angle) is needed to calculate the real flow ve- locity. Currently, the Doppler angle is determined visually by manually aligning a vessel axis marker along the blood vessel on the duplex scan image of the ultrasound. The ap- plication of this procedure is often limited by practical con- straints; therefore, measurements are not reliable. In order to overcome this problem, we developed a simple Doppler angle and flow velocity estimation method using a combi- nation of the classic and transverse Doppler effects. This method uses only a single focused annular array transducer to estimate the Doppler angle and the flow velocity. We have verified experimentally that this method is successful for measuring constant flow in a flow phantom between 45O and 80' Doppler angle. The standard deviation of the esti- mated Doppler angles is less than 4.5O. This method could be implemented easily in medical Doppler ultrasound sys- tems to automatically estimate the Doppler angle and the flow velocity.
Key concepts: Doppler effect, Flow velocity, Acoustic Doppler velocimetry, Imaging phantom, Flow (mathematics), Transverse plane, Ultrasonic flow meter, Transducer