Doppler angle and flow velocity estimations using the classic and transverse Doppler effects

Bor-Ray Lee, Huihua Kenny Chiang, Cheng‐Deng Kuo, Win-Li Lin, San-Kan Lee

Open publisher page 18 citations

Abstract

Current clinical Doppler ultrasound systems could only measure the flow vector parallel to the ultrasound beam axis, and the knowledge of the Doppler angle (beam-to-flow angle) is needed to calculate the real flow velocity. Currently, the Doppler angle is determined visually by manually aligning a vessel axis marker along the blood vessel on the duplex scan image of the ultrasound. The application of this procedure is often limited by practical constraints; therefore, measurements are not reliable. In order to overcome this problem, the authors developed a simple Doppler angle and flow velocity estimation method using a combination of the classic and transverse Doppler effects. This method uses only a single focused annular array transducer to estimate the Doppler angle and the flow velocity. The authors have verified experimentally that this method is successful for measuring constant flow in a flow phantom between 45 degrees and 80 degrees Doppler angle. The standard deviation of the estimated Doppler angles is less than 4.5 degrees . This method could be implemented easily in medical Doppler ultrasound systems to automatically estimate the Doppler angle and the flow velocity.

About this research paper

What this paper is about

Current clinical Doppler ultrasound systems could only measure the flow vector parallel to the ultrasound beam axis, and the knowledge of the Doppler angle (beam-to-flow angle) is needed to calculate the real flow velocity. Currently, the Doppler angle is determined visually by manually aligning a vessel axis marker along the blood vessel on the duplex scan image of the ultrasound. The application of this procedure is often limited by practical constraints; therefore, measurements are not reliable. In order to overcome this problem, the authors developed a simple Doppler angle and flow velocity estimation method using a combination of the classic and transverse Doppler effects. This method uses only a single focused annular array transducer to estimate the Doppler angle and the flow velocity. The authors have verified experimentally that this method is successful for measuring constant flow in a flow phantom between 45 degrees and 80 degrees Doppler angle. The standard deviation of the estimated Doppler angles is less than 4.5 degrees . This method could be implemented easily in medical Doppler ultrasound systems to automatically estimate the Doppler angle and the flow velocity.

Why it matters

OpenAlex reports 18 citations for this work. Citation counts describe recorded attention and do not establish research quality.

Key contribution

A contribution statement is not available in the OpenAlex record.

Method / approach

Method details are not available in the OpenAlex metadata.

Main findings

Findings are not separately available in the OpenAlex metadata.

Limitations

Limitations are not available in the OpenAlex metadata.

Applications

Application details are not available in the OpenAlex metadata.

Available abstract

Current clinical Doppler ultrasound systems could only measure the flow vector parallel to the ultrasound beam axis, and the knowledge of the Doppler angle (beam-to-flow angle) is needed to calculate the real flow velocity. Currently, the Doppler angle is determined visually by manually aligning a vessel axis marker along the blood vessel on the duplex scan image of the ultrasound. The application of this procedure is often limited by practical constraints; therefore, measurements are not reliable. In order to overcome this problem, the authors developed a simple Doppler angle and flow velocity estimation method using a combination of the classic and transverse Doppler effects. This method uses only a single focused annular array transducer to estimate the Doppler angle and the flow velocity. The authors have verified experimentally that this method is successful for measuring constant flow in a flow phantom between 45 degrees and 80 degrees Doppler angle. The standard deviation of the estimated Doppler angles is less than 4.5 degrees . This method could be implemented easily in medical Doppler ultrasound systems to automatically estimate the Doppler angle and the flow velocity.

Key concepts: Doppler effect, Acoustic Doppler velocimetry, Imaging phantom, Flow velocity, Transverse plane, Flow (mathematics), Ultrasonic flow meter, Optics