Measurements of central corneal thickness using two immersion ultrasound techniques and optical technique.

Haya Alfarhan

Open publisher page 1 citations

Abstract

OBJECTIVE: To compare the accuracy of central corneal thickness measurements using ultrasound biomicroscopy, Orbscan II tomography and an Artemis-2 very high frequency ultrasound scanner. METHODS: The prospective study was conducted at Eye World Centre, Riyadh, Saudi Arabia, from September to November 2012. One eye from each of 60 normal subjects was analysed. The central corneal thickness was measured using ultrasound biomicroscopy, Orbscan II tomography and the Artemis-2 very high frequency ultrasound scanner. Results were compared using analysis of variance, repeated-measures analysis of variance and limits of agreement. RESULTS: The mean central corneal thickness was 530.30 +/- 30.75 microm, 548.95 +/- 30.33 microm and 554.73 +/- 31.97 microm for biomicroscopy, tomography and the scanner respectively. The intraobserver repeatability analyses of variance were not significant for the three procedures (p = 0.19, 0.23 and 0.41, respectively). A significant difference was noted among the three different methods (p = 0.0001). However, comparison among instruments revealed no significant difference between tomography and the scanner (p > 0.05), yet significant differences were noted in biomicroscopy vs. tomography, and biomicroscopy vs. the scanner (p < 0.01 and P < 0.001, respectively). The mean differences (and upper/lower limits of agreement) for central corneal thickness measurements were 18.92 +/- 40.71 (60.90/-98.70); 24.7 +/- 13.13 (1.00/-50.40), and -5.80 +/- 38.61 (69.90/-81.40) for biomicroscopy vs. tomography, biomicroscopy vs. scanner, and tomography vs. scanner respectively. CONCLUSIONS: The central corneal thickenss measurements obtained using Orbscan II tomography and the Artemis-2 very high frequency ultrasound scanner can be used interchangeably. However, Orbscan II tomography and the Artemis-2 scanner measurements cannot be used interchangeably with ultrasound biomicroscopy.

About this research paper

What this paper is about

OBJECTIVE: To compare the accuracy of central corneal thickness measurements using ultrasound biomicroscopy, Orbscan II tomography and an Artemis-2 very high frequency ultrasound scanner. METHODS: The prospective study was conducted at Eye World Centre, Riyadh, Saudi Arabia, from September to November 2012. One eye from each of 60 normal subjects was analysed. The central corneal thickness was measured using ultrasound biomicroscopy, Orbscan II tomography and the Artemis-2 very high frequency ultrasound scanner. Results were compared using analysis of variance, repeated-measures analysis of variance and limits of agreement. RESULTS: The mean central corneal thickness was 530.30 +/- 30.75 microm, 548.95 +/- 30.33 microm and 554.73 +/- 31.97 microm for biomicroscopy, tomography and the scanner respectively. The intraobserver repeatability analyses of variance were not significant for the three procedures (p = 0.19, 0.23 and 0.41, respectively). A significant difference was noted among the three different methods (p = 0.0001). However, comparison among instruments revealed no significant difference between tomography and the scanner (p > 0.05), yet significant differences were noted in biomicroscopy vs. tomography, and biomicroscopy vs. the scanner (p < 0.01 and P < 0.001, respectively). The mean differences (and upper/lower limits of agreement) for central corneal thickness measurements were 18.92 +/- 40.71 (60.90/-98.70); 24.7 +/- 13.13 (1.00/-50.40), and -5.80 +/- 38.61 (69.90/-81.40) for biomicroscopy vs. tomography, biomicroscopy vs. scanner, and tomography vs. scanner respectively. CONCLUSIONS: The central corneal thickenss measurements obtained using Orbscan II tomography and the Artemis-2 very high frequency ultrasound scanner can be used interchangeably. However, Orbscan II tomography and the Artemis-2 scanner measurements cannot be used interchangeably with ultrasound biomicroscopy.

Why it matters

OpenAlex reports 1 citations for this work. Citation counts describe recorded attention and do not establish research quality.

Key contribution

A contribution statement is not available in the OpenAlex record.

Method / approach

Method details are not available in the OpenAlex metadata.

Main findings

Findings are not separately available in the OpenAlex metadata.

Limitations

Limitations are not available in the OpenAlex metadata.

Applications

Application details are not available in the OpenAlex metadata.

Available abstract

OBJECTIVE: To compare the accuracy of central corneal thickness measurements using ultrasound biomicroscopy, Orbscan II tomography and an Artemis-2 very high frequency ultrasound scanner. METHODS: The prospective study was conducted at Eye World Centre, Riyadh, Saudi Arabia, from September to November 2012. One eye from each of 60 normal subjects was analysed. The central corneal thickness was measured using ultrasound biomicroscopy, Orbscan II tomography and the Artemis-2 very high frequency ultrasound scanner. Results were compared using analysis of variance, repeated-measures analysis of variance and limits of agreement. RESULTS: The mean central corneal thickness was 530.30 +/- 30.75 microm, 548.95 +/- 30.33 microm and 554.73 +/- 31.97 microm for biomicroscopy, tomography and the scanner respectively. The intraobserver repeatability analyses of variance were not significant for the three procedures (p = 0.19, 0.23 and 0.41, respectively). A significant difference was noted among the three different methods (p = 0.0001). However, comparison among instruments revealed no significant difference between tomography and the scanner (p > 0.05), yet significant differences were noted in biomicroscopy vs. tomography, and biomicroscopy vs. the scanner (p < 0.01 and P < 0.001, respectively). The mean differences (and upper/lower limits of agreement) for central corneal thickness measurements were 18.92 +/- 40.71 (60.90/-98.70); 24.7 +/- 13.13 (1.00/-50.40), and -5.80 +/- 38.61 (69.90/-81.40) for biomicroscopy vs. tomography, biomicroscopy vs. scanner, and tomography vs. scanner respectively. CONCLUSIONS: The central corneal thickenss measurements obtained using Orbscan II tomography and the Artemis-2 very high frequency ultrasound scanner can be used interchangeably. However, Orbscan II tomography and the Artemis-2 scanner measurements cannot be used interchangeably with ultrasound biomicroscopy.

Key concepts: Ultrasound biomicroscopy, Medicine, Repeatability, Ophthalmology, Tomography, Ultrasound, Significant difference, Nuclear medicine

Related papers

Back to paper searchBrowse research topicsOriginal source
Measurements of central corneal thickness using two immersion ultrasound techniques and optical technique. — Research Paper | ScholarLens