Intra‐ and interobserver reliability of lens equatorial length measurement using 35‐ MH z ultrasound biomicroscopy in dogs with cataract

C. Barbé, Nathaniel X. Harran, Frédéric Goulle

Open publisher page 5 citations

Abstract

PURPOSE: To determine intra- and interobserver reliability of lens equatorial length measurement using 35-MHz ultrasound biomicroscopy in dogs with cataract. METHODS: Ocular ultrasonography was performed on 28 dogs (50 eyes) before phacoemulsification. Lens equatorial length was measured on still images obtained at 35 MHz in all dogs. Each observer (A and B) obtained the measurements independently. Coefficients of variation (CVs) were used to assess intraobserver repeatability and interobserver reproducibility. Intraobserver repeatability was assessed by comparing measurements obtained on the same image on four different occasions by observer A. Interobserver reproducibility was assessed by comparing measurements obtained independently by both observers on the same image. RESULTS: CV used to assess intraobserver repeatability was 2.4% with good agreement and no significant difference between repetitions (P > 0.05). CV used to assess interobserver reproducibility was 5.8% without significant difference between measurements obtained by the two observers (P > 0.05). CONCLUSIONS: Intraobserver repeatability and interobserver reproducibility were considered good for lens equatorial length measurement at 35 MHz in dogs with cataract. Ultrasound biomicroscopy could be an acceptable method to assess lens equatorial dimension before phacoemulsification and intraocular lens implantation in dogs with cataract.

About this research paper

What this paper is about

PURPOSE: To determine intra- and interobserver reliability of lens equatorial length measurement using 35-MHz ultrasound biomicroscopy in dogs with cataract. METHODS: Ocular ultrasonography was performed on 28 dogs (50 eyes) before phacoemulsification. Lens equatorial length was measured on still images obtained at 35 MHz in all dogs. Each observer (A and B) obtained the measurements independently. Coefficients of variation (CVs) were used to assess intraobserver repeatability and interobserver reproducibility. Intraobserver repeatability was assessed by comparing measurements obtained on the same image on four different occasions by observer A. Interobserver reproducibility was assessed by comparing measurements obtained independently by both observers on the same image. RESULTS: CV used to assess intraobserver repeatability was 2.4% with good agreement and no significant difference between repetitions (P > 0.05). CV used to assess interobserver reproducibility was 5.8% without significant difference between measurements obtained by the two observers (P > 0.05). CONCLUSIONS: Intraobserver repeatability and interobserver reproducibility were considered good for lens equatorial length measurement at 35 MHz in dogs with cataract. Ultrasound biomicroscopy could be an acceptable method to assess lens equatorial dimension before phacoemulsification and intraocular lens implantation in dogs with cataract.

Why it matters

OpenAlex reports 5 citations for this work. Citation counts describe recorded attention and do not establish research quality.

Key contribution

A contribution statement is not available in the OpenAlex record.

Method / approach

Method details are not available in the OpenAlex metadata.

Main findings

Findings are not separately available in the OpenAlex metadata.

Limitations

Limitations are not available in the OpenAlex metadata.

Applications

Application details are not available in the OpenAlex metadata.

Available abstract

PURPOSE: To determine intra- and interobserver reliability of lens equatorial length measurement using 35-MHz ultrasound biomicroscopy in dogs with cataract. METHODS: Ocular ultrasonography was performed on 28 dogs (50 eyes) before phacoemulsification. Lens equatorial length was measured on still images obtained at 35 MHz in all dogs. Each observer (A and B) obtained the measurements independently. Coefficients of variation (CVs) were used to assess intraobserver repeatability and interobserver reproducibility. Intraobserver repeatability was assessed by comparing measurements obtained on the same image on four different occasions by observer A. Interobserver reproducibility was assessed by comparing measurements obtained independently by both observers on the same image. RESULTS: CV used to assess intraobserver repeatability was 2.4% with good agreement and no significant difference between repetitions (P > 0.05). CV used to assess interobserver reproducibility was 5.8% without significant difference between measurements obtained by the two observers (P > 0.05). CONCLUSIONS: Intraobserver repeatability and interobserver reproducibility were considered good for lens equatorial length measurement at 35 MHz in dogs with cataract. Ultrasound biomicroscopy could be an acceptable method to assess lens equatorial dimension before phacoemulsification and intraocular lens implantation in dogs with cataract.

Key concepts: Repeatability, Reproducibility, Ultrasound biomicroscopy, Medicine, Ultrasound, Phacoemulsification, Ophthalmology, Nuclear medicine