Recruitment data for nerve cuff electrodes: implications for design of implantable stimulators

Donald R. McNeal, Lucinda L. Baker, J.T. Symons

Open publisher page 35 citations

Abstract

Recruitment characteristics of nerve cuff electrodes implanted in four cats for five months were measured. Monopolar, bipolar, and tripolar configurations were considered. Approximately twice the current was required to achieve a given response using the tripolar configuration as compared with monopolar stimulation. Bipolar stimulation also required more current than monopolar stimulation. Using the recruitment data, a number of strategies for modulating muscle tension were considered. It was shown that both pulse amplitude and pulse duration should be software-selectable to achieve adequate control of muscle tension when using either pulse amplitude modulation or pulse duration modulation. When using pulse amplitude modulation, it was found to be desirable to operate at a low pulse duration in the high end of the allowable range for pulse amplitude. For pulse duration modulation, one should operate at a low pulse amplitude in the high end of the allowable range for pulse duration. The effect of pulse amplitude and pulse duration step size on the maximum step change in muscle tension and the linearity of the recruitment curves were examined. The use of logarithmic steps in the modulation parameter was examined and was shown to result in improved controllability and linearity.

About this research paper

What this paper is about

Recruitment characteristics of nerve cuff electrodes implanted in four cats for five months were measured. Monopolar, bipolar, and tripolar configurations were considered. Approximately twice the current was required to achieve a given response using the tripolar configuration as compared with monopolar stimulation. Bipolar stimulation also required more current than monopolar stimulation. Using the recruitment data, a number of strategies for modulating muscle tension were considered. It was shown that both pulse amplitude and pulse duration should be software-selectable to achieve adequate control of muscle tension when using either pulse amplitude modulation or pulse duration modulation. When using pulse amplitude modulation, it was found to be desirable to operate at a low pulse duration in the high end of the allowable range for pulse amplitude. For pulse duration modulation, one should operate at a low pulse amplitude in the high end of the allowable range for pulse duration. The effect of pulse amplitude and pulse duration step size on the maximum step change in muscle tension and the linearity of the recruitment curves were examined. The use of logarithmic steps in the modulation parameter was examined and was shown to result in improved controllability and linearity.

Why it matters

OpenAlex reports 35 citations for this work. Citation counts describe recorded attention and do not establish research quality.

Key contribution

A contribution statement is not available in the OpenAlex record.

Method / approach

Method details are not available in the OpenAlex metadata.

Main findings

Findings are not separately available in the OpenAlex metadata.

Limitations

Limitations are not available in the OpenAlex metadata.

Applications

Application details are not available in the OpenAlex metadata.

Available abstract

Recruitment characteristics of nerve cuff electrodes implanted in four cats for five months were measured. Monopolar, bipolar, and tripolar configurations were considered. Approximately twice the current was required to achieve a given response using the tripolar configuration as compared with monopolar stimulation. Bipolar stimulation also required more current than monopolar stimulation. Using the recruitment data, a number of strategies for modulating muscle tension were considered. It was shown that both pulse amplitude and pulse duration should be software-selectable to achieve adequate control of muscle tension when using either pulse amplitude modulation or pulse duration modulation. When using pulse amplitude modulation, it was found to be desirable to operate at a low pulse duration in the high end of the allowable range for pulse amplitude. For pulse duration modulation, one should operate at a low pulse amplitude in the high end of the allowable range for pulse duration. The effect of pulse amplitude and pulse duration step size on the maximum step change in muscle tension and the linearity of the recruitment curves were examined. The use of logarithmic steps in the modulation parameter was examined and was shown to result in improved controllability and linearity.

Key concepts: Pulse-amplitude modulation, Pulse duration, Pulse (music), Pulse-width modulation, Amplitude, Materials science, Amplitude modulation, Biomedical engineering

Related papers

Back to paper searchBrowse research topicsOriginal source
Recruitment data for nerve cuff electrodes: implications for design of implantable stimulators — Research Paper | ScholarLens