Optimization of Criteria for Verification of Automated Platelet Counts Generated by the Sysmex XE-2100 Hematology Analyzer
Gene Gulati, Sean Cote, Eric Behling, William Kocher
Abstract
Gene Gulati, Sean Cote, Eric Behling, William Kocher
Abstract
Background Verification of automated platelet counts by blood smear examination is necessary but time consuming. This study was undertaken to optimize the criteria for performing platelet scans. Methods Platelet scans were performed on 796 blood smears made from specimens that were either flagged for (a) platelet abnormal distribution or PAD (group I), (b) platelet clumps or CLP (group II), (c) PAD + CLP (group III), or (d) had a platelet count below 100,000/μL (group IV). Based on the findings of the platelet scans, a percent positive yield was determined for each group of specimens. Results Percent positive yields of 8.9, 30.0, 35.3, and 1.9 were obtained from platelet scan findings for groups I, II, III, and IV, respectively. Conclusions Optimal criteria for performing platelet scans excluded the PAD flag and the follow-up platelet counts below 100,000/μL but included specimens flagged for CLP and the initial platelet counts below 100,000/μL.
OpenAlex reports 4 citations for this work. Citation counts describe recorded attention and do not establish research quality.
A contribution statement is not available in the OpenAlex record.
Method details are not available in the OpenAlex metadata.
Findings are not separately available in the OpenAlex metadata.
Limitations are not available in the OpenAlex metadata.
Application details are not available in the OpenAlex metadata.
Background Verification of automated platelet counts by blood smear examination is necessary but time consuming. This study was undertaken to optimize the criteria for performing platelet scans. Methods Platelet scans were performed on 796 blood smears made from specimens that were either flagged for (a) platelet abnormal distribution or PAD (group I), (b) platelet clumps or CLP (group II), (c) PAD + CLP (group III), or (d) had a platelet count below 100,000/μL (group IV). Based on the findings of the platelet scans, a percent positive yield was determined for each group of specimens. Results Percent positive yields of 8.9, 30.0, 35.3, and 1.9 were obtained from platelet scan findings for groups I, II, III, and IV, respectively. Conclusions Optimal criteria for performing platelet scans excluded the PAD flag and the follow-up platelet counts below 100,000/μL but included specimens flagged for CLP and the initial platelet counts below 100,000/μL.
Key concepts: Hematology analyzer, Spectrum analyzer, Hematology, Medicine, Medical physics, Internal medicine, Computer science, Telecommunications