Electrical detection of the contact between a microinjection pipette and cells

Mikko Lukkari, M.I. Karjalainen, R. Sarkanen, Marja‐Leena Linne, Tuula O. Jalonen, Pasi Kallio

Open publisher page 13 citations

Abstract

This work presents the basic design and tests of a device designed for detecting the contact between a microinjection pipette and cell membrane. The device facilitates the automation of the microinjection procedure of living adherent cells. The measurement of the contact is based on measuring the resistance of the pipette. Breakage and clogging of the pipette can be detected with the same measurement.

About this research paper

What this paper is about

This work presents the basic design and tests of a device designed for detecting the contact between a microinjection pipette and cell membrane. The device facilitates the automation of the microinjection procedure of living adherent cells. The measurement of the contact is based on measuring the resistance of the pipette. Breakage and clogging of the pipette can be detected with the same measurement.

Why it matters

OpenAlex reports 13 citations for this work. Citation counts describe recorded attention and do not establish research quality.

Key contribution

A contribution statement is not available in the OpenAlex record.

Method / approach

Method details are not available in the OpenAlex metadata.

Main findings

Findings are not separately available in the OpenAlex metadata.

Limitations

Limitations are not available in the OpenAlex metadata.

Applications

Application details are not available in the OpenAlex metadata.

Available abstract

This work presents the basic design and tests of a device designed for detecting the contact between a microinjection pipette and cell membrane. The device facilitates the automation of the microinjection procedure of living adherent cells. The measurement of the contact is based on measuring the resistance of the pipette. Breakage and clogging of the pipette can be detected with the same measurement.

Key concepts: Pipette, Microinjection, Biomedical engineering, Clogging, Materials science, Chemistry, Engineering, Biology

Related papers

Back to paper searchBrowse research topicsOriginal source
Electrical detection of the contact between a microinjection pipette and cells — Research Paper | ScholarLens