Performance evaluation of the Sysmex XN ‐1000 hematology analyzer in assessment of the white blood cell count differential in pediatric specimens
P.‐H. Becker, Odile Fenneteau, Lydie Da Costa
Abstract
P.‐H. Becker, Odile Fenneteau, Lydie Da Costa
Abstract
INTRODUCTION: The automated XN-1000 hematology analyzer enables to perform a blood cell count and a leukocyte differential. When abnormal cells were detected, a flag was generated by the analyzer and a manual microscopic examination of the corresponding blood film was performed. METHODS: We compared the white blood cell differentials provided by the automated hematology analyzer XN-1000 in a pediatric population (n = 765) with those obtained through microscopic examination by cytologists and those obtained using a previous version of this analyzer, the XE-2100. Leukocytes count as well as flags sensitivity and specificity was analyzed. RESULTS: The leukocytes count provided by the analyzer is in good accordance with the differential obtained by manual count in children older than 3 months. The sensitivity for blast detection is 99% and the detection of reactive cells is 63%. The flag specificity remains low (<35%) for blood samples collected from infants between 8 days and 2 years of age, but increases up to 67% thereafter. The results obtained with the XN-1000 analyzer show an improvement in comparison with those obtained with the XE-2100 analyzer. CONCLUSION: The automated WBC differential provided by the XN-1000 analyzer in the pediatric setting is accurate, but a meticulous microscopic examination of blood smears remains necessary for infants up to 3 months of age to validate the analyzer flags.
OpenAlex reports 31 citations for this work. Citation counts describe recorded attention and do not establish research quality.
A contribution statement is not available in the OpenAlex record.
Method details are not available in the OpenAlex metadata.
Findings are not separately available in the OpenAlex metadata.
Limitations are not available in the OpenAlex metadata.
Application details are not available in the OpenAlex metadata.
INTRODUCTION: The automated XN-1000 hematology analyzer enables to perform a blood cell count and a leukocyte differential. When abnormal cells were detected, a flag was generated by the analyzer and a manual microscopic examination of the corresponding blood film was performed. METHODS: We compared the white blood cell differentials provided by the automated hematology analyzer XN-1000 in a pediatric population (n = 765) with those obtained through microscopic examination by cytologists and those obtained using a previous version of this analyzer, the XE-2100. Leukocytes count as well as flags sensitivity and specificity was analyzed. RESULTS: The leukocytes count provided by the analyzer is in good accordance with the differential obtained by manual count in children older than 3 months. The sensitivity for blast detection is 99% and the detection of reactive cells is 63%. The flag specificity remains low (<35%) for blood samples collected from infants between 8 days and 2 years of age, but increases up to 67% thereafter. The results obtained with the XN-1000 analyzer show an improvement in comparison with those obtained with the XE-2100 analyzer. CONCLUSION: The automated WBC differential provided by the XN-1000 analyzer in the pediatric setting is accurate, but a meticulous microscopic examination of blood smears remains necessary for infants up to 3 months of age to validate the analyzer flags.
Key concepts: Hematology analyzer, Spectrum analyzer, White blood cell, Medicine, Hematology, Blood cell, Pathology, Population