2012•Review of Scientific InstrumentsOpen access

Polarization pyrometry: An improvement to multi-wavelength optical pyrometry

P. Ni, Richard M. More, Hitoki Yoneda, Frank M. Bieniosek

Open full text 14 citations

Abstract

We describe a new method that improves upon temperature measurement by optical pyrometry. The main uncertainty in the traditional pyrometry technique is the surface emissivity, which is generally unknown and hard to measure. A common approach to deal with this problem is to measure the thermal emission at multiple wavelengths - an approach called multi-wavelength pyrometry. However, this technique can still result in a level of uncertainty in the surface temperature that is unsatisfactory for scientific applications, such as a measurement of equation of state of warm dense matter. In contrast to the conventional multi-wavelength technique, in the polarization pyrometry approach described herein, p- and s-polarization components of thermal radiation at multiple-angles are used to deduce the temperature. This paper describes the concept and the results of an initial proof-of-principle static experiment with an electrically heated tungsten ribbon. It was found that in the same experiment, the accuracy of the polarization pyrometry measurement was substantially greater than that achieved using conventional multi-wavelength pyrometry.

Open-access reader

About this research paper

What this paper is about

We describe a new method that improves upon temperature measurement by optical pyrometry. The main uncertainty in the traditional pyrometry technique is the surface emissivity, which is generally unknown and hard to measure. A common approach to deal with this problem is to measure the thermal emission at multiple wavelengths - an approach called multi-wavelength pyrometry. However, this technique can still result in a level of uncertainty in the surface temperature that is unsatisfactory for scientific applications, such as a measurement of equation of state of warm dense matter. In contrast to the conventional multi-wavelength technique, in the polarization pyrometry approach described herein, p- and s-polarization components of thermal radiation at multiple-angles are used to deduce the temperature. This paper describes the concept and the results of an initial proof-of-principle static experiment with an electrically heated tungsten ribbon. It was found that in the same experiment, the accuracy of the polarization pyrometry measurement was substantially greater than that achieved using conventional multi-wavelength pyrometry.

Why it matters

OpenAlex reports 14 citations for this work. Citation counts describe recorded attention and do not establish research quality.

Key contribution

A contribution statement is not available in the OpenAlex record.

Method / approach

Method details are not available in the OpenAlex metadata.

Main findings

Findings are not separately available in the OpenAlex metadata.

Limitations

Limitations are not available in the OpenAlex metadata.

Applications

Application details are not available in the OpenAlex metadata.

Available abstract

We describe a new method that improves upon temperature measurement by optical pyrometry. The main uncertainty in the traditional pyrometry technique is the surface emissivity, which is generally unknown and hard to measure. A common approach to deal with this problem is to measure the thermal emission at multiple wavelengths - an approach called multi-wavelength pyrometry. However, this technique can still result in a level of uncertainty in the surface temperature that is unsatisfactory for scientific applications, such as a measurement of equation of state of warm dense matter. In contrast to the conventional multi-wavelength technique, in the polarization pyrometry approach described herein, p- and s-polarization components of thermal radiation at multiple-angles are used to deduce the temperature. This paper describes the concept and the results of an initial proof-of-principle static experiment with an electrically heated tungsten ribbon. It was found that in the same experiment, the accuracy of the polarization pyrometry measurement was substantially greater than that achieved using conventional multi-wavelength pyrometry.

Key concepts: Pyrometer, Emissivity, Wavelength, Optics, Polarization (electrochemistry), Temperature measurement, Materials science, Radiation

Related papers