Corneal thickness, curvature, and elevation readings in normal corneas: Combined Placido–Scheimpflug system versus combined Placido–scanning-slit system

Emmanuel Guilbert, Alain Saad, Alice Grise-Dulac, Damien Gatinel

Open publisher page 49 citations

Abstract

PURPOSE: To evaluate agreement in central corneal thickness (CCT), keratometry, and anterior and posterior elevation map measurements in normal corneas between a combined Placido-Scheimpflug system and a combined Placido-scanning-slit elevation topography system. SETTING: Department of Cataract & Refractive Surgery, Rothschild Foundation, Paris, France. DESIGN: Evaluation of diagnostic test or technology. METHODS: Measurements were performed with a combined Placido-Scheimpflug system (TMS-5) and a combined Placido-scanning-slit system (Orbscan II). Ultrasound (US) pachymetry was used as the reference for CCT measurements. Bland-Altman plots were used to evaluate agreement between instruments. RESULTS: The mean CCT measurements by US pachymetry, the Placido-Scheimpflug system, and the Placido-scanning-slit system were 556.74 μm ± 42.45 (SD), 543.23 ± 36.73 μm, and 564.45 ± 41.26 μm, respectively. Although the CCT readings were statistically significantly thinner with the Placido-Scheimpflug system than with the other systems, there was high correlation between instruments. Peripheral corneal thickness readings were also thinner with the Placido-Scheimpflug system than with the Placido-scanning-slit system. Keratometry and anterior and posterior best-fit sphere (BFS) measurements were comparable between the 2 optical devices. Anterior and posterior maximum central elevations measured by the 2 instruments were not comparable or strongly correlated. Repeatability after 3 successive measurements was excellent for all parameters except maximum central elevation. CONCLUSIONS: Although highly correlated, with corneal thickness readings were not interchangeable between the 2 optical devices. No statistically significant differences in keratometry or BFS measurements were observed between the 2 devices. There were important discrepancies in the maximum central elevation between the 2 topographers. FINANCIAL DISCLOSURE: Drs. Gatinel and Saad are consultants to Technolas Perfect Vision. No other author has a financial or proprietary interest in any material or method mentioned.

About this research paper

What this paper is about

PURPOSE: To evaluate agreement in central corneal thickness (CCT), keratometry, and anterior and posterior elevation map measurements in normal corneas between a combined Placido-Scheimpflug system and a combined Placido-scanning-slit elevation topography system. SETTING: Department of Cataract & Refractive Surgery, Rothschild Foundation, Paris, France. DESIGN: Evaluation of diagnostic test or technology. METHODS: Measurements were performed with a combined Placido-Scheimpflug system (TMS-5) and a combined Placido-scanning-slit system (Orbscan II). Ultrasound (US) pachymetry was used as the reference for CCT measurements. Bland-Altman plots were used to evaluate agreement between instruments. RESULTS: The mean CCT measurements by US pachymetry, the Placido-Scheimpflug system, and the Placido-scanning-slit system were 556.74 μm ± 42.45 (SD), 543.23 ± 36.73 μm, and 564.45 ± 41.26 μm, respectively. Although the CCT readings were statistically significantly thinner with the Placido-Scheimpflug system than with the other systems, there was high correlation between instruments. Peripheral corneal thickness readings were also thinner with the Placido-Scheimpflug system than with the Placido-scanning-slit system. Keratometry and anterior and posterior best-fit sphere (BFS) measurements were comparable between the 2 optical devices. Anterior and posterior maximum central elevations measured by the 2 instruments were not comparable or strongly correlated. Repeatability after 3 successive measurements was excellent for all parameters except maximum central elevation. CONCLUSIONS: Although highly correlated, with corneal thickness readings were not interchangeable between the 2 optical devices. No statistically significant differences in keratometry or BFS measurements were observed between the 2 devices. There were important discrepancies in the maximum central elevation between the 2 topographers. FINANCIAL DISCLOSURE: Drs. Gatinel and Saad are consultants to Technolas Perfect Vision. No other author has a financial or proprietary interest in any material or method mentioned.

Why it matters

OpenAlex reports 49 citations for this work. Citation counts describe recorded attention and do not establish research quality.

Key contribution

A contribution statement is not available in the OpenAlex record.

Method / approach

Method details are not available in the OpenAlex metadata.

Main findings

Findings are not separately available in the OpenAlex metadata.

Limitations

Limitations are not available in the OpenAlex metadata.

Applications

Application details are not available in the OpenAlex metadata.

Available abstract

PURPOSE: To evaluate agreement in central corneal thickness (CCT), keratometry, and anterior and posterior elevation map measurements in normal corneas between a combined Placido-Scheimpflug system and a combined Placido-scanning-slit elevation topography system. SETTING: Department of Cataract & Refractive Surgery, Rothschild Foundation, Paris, France. DESIGN: Evaluation of diagnostic test or technology. METHODS: Measurements were performed with a combined Placido-Scheimpflug system (TMS-5) and a combined Placido-scanning-slit system (Orbscan II). Ultrasound (US) pachymetry was used as the reference for CCT measurements. Bland-Altman plots were used to evaluate agreement between instruments. RESULTS: The mean CCT measurements by US pachymetry, the Placido-Scheimpflug system, and the Placido-scanning-slit system were 556.74 μm ± 42.45 (SD), 543.23 ± 36.73 μm, and 564.45 ± 41.26 μm, respectively. Although the CCT readings were statistically significantly thinner with the Placido-Scheimpflug system than with the other systems, there was high correlation between instruments. Peripheral corneal thickness readings were also thinner with the Placido-Scheimpflug system than with the Placido-scanning-slit system. Keratometry and anterior and posterior best-fit sphere (BFS) measurements were comparable between the 2 optical devices. Anterior and posterior maximum central elevations measured by the 2 instruments were not comparable or strongly correlated. Repeatability after 3 successive measurements was excellent for all parameters except maximum central elevation. CONCLUSIONS: Although highly correlated, with corneal thickness readings were not interchangeable between the 2 optical devices. No statistically significant differences in keratometry or BFS measurements were observed between the 2 devices. There were important discrepancies in the maximum central elevation between the 2 topographers. FINANCIAL DISCLOSURE: Drs. Gatinel and Saad are consultants to Technolas Perfect Vision. No other author has a financial or proprietary interest in any material or method mentioned.

Key concepts: Scheimpflug principle, Keratometer, Corneal topography, Optics, Ophthalmology, Cornea, Repeatability, Refractive surgery