Repeatability and Inter-device Agreement for Three Different Methods of Keratometry: Placido, Scheimpflug, and Color LED Corneal Topography

Irene Ruiz Hidalgo, J. Rozema, Sorcha Ní Dhubhghaill, Nadia Zakaria, Carina Koppen, Marie‐José Tassignon

Open publisher page 36 citations

Abstract

PURPOSE: To determine the repeatability of a color LED corneal topographer (Cassini; iOptics, The Hague, The Netherlands) and compare it with Placido and Scheimpflug based devices (EyeSys 2000; EyeSys Laboratories, Houston, TX, and Pentacam HR; Oculus Optikgeräte GmbH, Wetzlar, Germany). METHODS: This prospective study involved 20 healthy volunteers (20 eyes) recruited from the staff of the Antwerp University Hospital. For each eye, three measurements were taken using each device, from which eight parameters describing keratometry and astigmatism were derived. Repeatability was assessed using intrasession, within-subject analysis of variance. Agreement was evaluated between pairs of devices with Bland- Altman plots and 95% confidence intervals. RESULTS: The repeatability values were good for all three devices, although slightly worse for the Cassini than for the other two devices for steep, flat, and mean keratometry. The EyeSys showed worse repeatability for the astigmatism axis. EyeSys and Pentacam agreed well with each other in terms of mean keratometry, whereas the Cassini gave consistently higher values by 0.52 diopters (D) when compared with the EyeSys (P < .05, paired t test) and by 0.38 D when compared with the Pentacam (P < .05, paired t test). The Cassini provided similar repeatability values (2.31°) for the astigmatism axis to the Pentacam (2.22°), in contrast to the EyeSys, which produced much more variable axes (9.0°) (P < .05, paired t test). CONCLUSIONS: Overall, the three devices display comparable repeatability. The one exception is the astigmatism axis measurement of the EyeSys. The Cassini provides higher keratometry values than the other two devices, but the astigmatism axis agrees well with that provided by the Pentacam.

About this research paper

What this paper is about

PURPOSE: To determine the repeatability of a color LED corneal topographer (Cassini; iOptics, The Hague, The Netherlands) and compare it with Placido and Scheimpflug based devices (EyeSys 2000; EyeSys Laboratories, Houston, TX, and Pentacam HR; Oculus Optikgeräte GmbH, Wetzlar, Germany). METHODS: This prospective study involved 20 healthy volunteers (20 eyes) recruited from the staff of the Antwerp University Hospital. For each eye, three measurements were taken using each device, from which eight parameters describing keratometry and astigmatism were derived. Repeatability was assessed using intrasession, within-subject analysis of variance. Agreement was evaluated between pairs of devices with Bland- Altman plots and 95% confidence intervals. RESULTS: The repeatability values were good for all three devices, although slightly worse for the Cassini than for the other two devices for steep, flat, and mean keratometry. The EyeSys showed worse repeatability for the astigmatism axis. EyeSys and Pentacam agreed well with each other in terms of mean keratometry, whereas the Cassini gave consistently higher values by 0.52 diopters (D) when compared with the EyeSys (P < .05, paired t test) and by 0.38 D when compared with the Pentacam (P < .05, paired t test). The Cassini provided similar repeatability values (2.31°) for the astigmatism axis to the Pentacam (2.22°), in contrast to the EyeSys, which produced much more variable axes (9.0°) (P < .05, paired t test). CONCLUSIONS: Overall, the three devices display comparable repeatability. The one exception is the astigmatism axis measurement of the EyeSys. The Cassini provides higher keratometry values than the other two devices, but the astigmatism axis agrees well with that provided by the Pentacam.

Why it matters

OpenAlex reports 36 citations for this work. Citation counts describe recorded attention and do not establish research quality.

Key contribution

A contribution statement is not available in the OpenAlex record.

Method / approach

Method details are not available in the OpenAlex metadata.

Main findings

Findings are not separately available in the OpenAlex metadata.

Limitations

Limitations are not available in the OpenAlex metadata.

Applications

Application details are not available in the OpenAlex metadata.

Available abstract

PURPOSE: To determine the repeatability of a color LED corneal topographer (Cassini; iOptics, The Hague, The Netherlands) and compare it with Placido and Scheimpflug based devices (EyeSys 2000; EyeSys Laboratories, Houston, TX, and Pentacam HR; Oculus Optikgeräte GmbH, Wetzlar, Germany). METHODS: This prospective study involved 20 healthy volunteers (20 eyes) recruited from the staff of the Antwerp University Hospital. For each eye, three measurements were taken using each device, from which eight parameters describing keratometry and astigmatism were derived. Repeatability was assessed using intrasession, within-subject analysis of variance. Agreement was evaluated between pairs of devices with Bland- Altman plots and 95% confidence intervals. RESULTS: The repeatability values were good for all three devices, although slightly worse for the Cassini than for the other two devices for steep, flat, and mean keratometry. The EyeSys showed worse repeatability for the astigmatism axis. EyeSys and Pentacam agreed well with each other in terms of mean keratometry, whereas the Cassini gave consistently higher values by 0.52 diopters (D) when compared with the EyeSys (P < .05, paired t test) and by 0.38 D when compared with the Pentacam (P < .05, paired t test). The Cassini provided similar repeatability values (2.31°) for the astigmatism axis to the Pentacam (2.22°), in contrast to the EyeSys, which produced much more variable axes (9.0°) (P < .05, paired t test). CONCLUSIONS: Overall, the three devices display comparable repeatability. The one exception is the astigmatism axis measurement of the EyeSys. The Cassini provides higher keratometry values than the other two devices, but the astigmatism axis agrees well with that provided by the Pentacam.

Key concepts: Repeatability, Keratometer, Scheimpflug principle, Astigmatism, Corneal topography, Dioptre, Ophthalmology, Optometry

Related papers

Back to paper searchBrowse research topicsOriginal source
Repeatability and Inter-device Agreement for Three Different Methods of Keratometry: Placido, Scheimpflug, and Color LED Corneal Topography — Research Paper | ScholarLens