Automated keratometry in routine cataract surgery: Comparison of Scheimpflug and conventional values
Richard J. Symes, Paul G. Ursell
Abstract
Richard J. Symes, Paul G. Ursell
Abstract
PURPOSE: To compare the mean keratometry (K) readings obtained with a conventional automated keratometer (IOLMaster) and a Scheimpflug keratometer (Pentacam) in eyes having preoperative assessment for routine cataract surgery. SETTING: Epsom and St. Helier University Hospitals, London, United Kingdom. DESIGN: Evaluation of diagnostic technology. METHODS: Mean K values were obtained with the conventional and Scheimpflug keratometers. The following Scheimpflug readings were evaluated: anterior K, true net power, and Holladay equivalent K measured at 1.0 to 7.0 mm corneal diameters. Mean readings for each type of keratometry were compared. Bland-Altman plots were used to determine the 95% limits of agreement (LoA) for the conventional and Scheimpflug keratometers. RESULTS: The mean conventional K was statistically significantly greater than the mean Scheimpflug K for true net power and equivalent K at 1.0 mm, 2.0 mm, 3.0 mm, and 4.0 mm corneal diameters. The mean conventional K was significantly less than the equivalent K at 5.0 mm, 6.0 mm, and 7.0 mm. The smallest mean difference was for equivalent K at 4.5 mm (0.02 diopters [D]). The smallest 95% LoA were -0.68 to 1.16 D (equivalent K at 5.0 mm) and -0.91 to 0.95 D (equivalent K at 4.5 mm). CONCLUSIONS: Overall, the equivalent K at 4.5 mm had the closest match with the conventional K values. The degree of interdevice variability with the conventional keratometer was lowest for the equivalent K at 4.5 mm and 5.0 mm, although this variability would be sufficient to influence intraocular lens power selection.
OpenAlex reports 51 citations for this work. Citation counts describe recorded attention and do not establish research quality.
A contribution statement is not available in the OpenAlex record.
Method details are not available in the OpenAlex metadata.
Findings are not separately available in the OpenAlex metadata.
Limitations are not available in the OpenAlex metadata.
Application details are not available in the OpenAlex metadata.
PURPOSE: To compare the mean keratometry (K) readings obtained with a conventional automated keratometer (IOLMaster) and a Scheimpflug keratometer (Pentacam) in eyes having preoperative assessment for routine cataract surgery. SETTING: Epsom and St. Helier University Hospitals, London, United Kingdom. DESIGN: Evaluation of diagnostic technology. METHODS: Mean K values were obtained with the conventional and Scheimpflug keratometers. The following Scheimpflug readings were evaluated: anterior K, true net power, and Holladay equivalent K measured at 1.0 to 7.0 mm corneal diameters. Mean readings for each type of keratometry were compared. Bland-Altman plots were used to determine the 95% limits of agreement (LoA) for the conventional and Scheimpflug keratometers. RESULTS: The mean conventional K was statistically significantly greater than the mean Scheimpflug K for true net power and equivalent K at 1.0 mm, 2.0 mm, 3.0 mm, and 4.0 mm corneal diameters. The mean conventional K was significantly less than the equivalent K at 5.0 mm, 6.0 mm, and 7.0 mm. The smallest mean difference was for equivalent K at 4.5 mm (0.02 diopters [D]). The smallest 95% LoA were -0.68 to 1.16 D (equivalent K at 5.0 mm) and -0.91 to 0.95 D (equivalent K at 4.5 mm). CONCLUSIONS: Overall, the equivalent K at 4.5 mm had the closest match with the conventional K values. The degree of interdevice variability with the conventional keratometer was lowest for the equivalent K at 4.5 mm and 5.0 mm, although this variability would be sufficient to influence intraocular lens power selection.
Key concepts: Scheimpflug principle, Keratometer, Dioptre, Cataract surgery, Ophthalmology, Limits of agreement, Medicine, Mathematics