Comparison of the oscillometric blood pressure monitor (BPM-100 Beta ) with the auscultatory mercury sphygmomanometer
Gurdial S. Mattu, Thomas L. Perry, James M Wright
Abstract
Gurdial S. Mattu, Thomas L. Perry, James M Wright
Abstract
BACKGROUND: To compare directly the accuracy of the BPM-100(Beta) monitor (an automated oscillometric blood pressure device) with standard auscultatory mercury sphygmomanometry. DESIGN: The BPM-100(Beta) was connected in parallel via a T-tube to a mercury sphygmomanometer. The BPM-100(Beta) and two trained observers (blinded from each other and from the BPM-100(Beta)) measured the sitting blood pressure simultaneously. METHODS: Means, standard deviations and ranges were calculated for all the demographic data: age, arm size, heart rate and blood pressure. The agreement between the BPM-100(Beta) and the mean of two observers (the reference) was determined and expressed as the mean +/- SD, as well as the percentage of differences falling within 5, 10 and 15 mmHg. RESULTS: Of the 92 subjects recruited, 85 (92.4%) met the inclusion criteria, and 391 sets of sitting blood pressure and heart rate measurements were available for analysis. The mean difference between the BPM-100(Beta) monitor and the reference was -0.62 +/- 6.96 mmHg for systolic blood pressure, -1.48 +/- 4.80 mmHg for diastolic blood pressure and 0.14 +/- 1.86 beats/min for heart rate. The only limitation of the device was its tendency to underestimate higher systolic blood pressures. This problem has been addressed by a minor change in the algorithm (see the companion publication, Blood Press Monit, 6, 161-165, 2001). CONCLUSION: The BPM-100(Beta) is an accurate blood pressure monitor for the office setting, meeting all requirements of the Association for the Advancement of Medical Instrumentation and achieving an 'A' grade according to the British Hypertension Society protocol.
OpenAlex reports 83 citations for this work. Citation counts describe recorded attention and do not establish research quality.
A contribution statement is not available in the OpenAlex record.
Method details are not available in the OpenAlex metadata.
Findings are not separately available in the OpenAlex metadata.
Limitations are not available in the OpenAlex metadata.
Application details are not available in the OpenAlex metadata.
BACKGROUND: To compare directly the accuracy of the BPM-100(Beta) monitor (an automated oscillometric blood pressure device) with standard auscultatory mercury sphygmomanometry. DESIGN: The BPM-100(Beta) was connected in parallel via a T-tube to a mercury sphygmomanometer. The BPM-100(Beta) and two trained observers (blinded from each other and from the BPM-100(Beta)) measured the sitting blood pressure simultaneously. METHODS: Means, standard deviations and ranges were calculated for all the demographic data: age, arm size, heart rate and blood pressure. The agreement between the BPM-100(Beta) and the mean of two observers (the reference) was determined and expressed as the mean +/- SD, as well as the percentage of differences falling within 5, 10 and 15 mmHg. RESULTS: Of the 92 subjects recruited, 85 (92.4%) met the inclusion criteria, and 391 sets of sitting blood pressure and heart rate measurements were available for analysis. The mean difference between the BPM-100(Beta) monitor and the reference was -0.62 +/- 6.96 mmHg for systolic blood pressure, -1.48 +/- 4.80 mmHg for diastolic blood pressure and 0.14 +/- 1.86 beats/min for heart rate. The only limitation of the device was its tendency to underestimate higher systolic blood pressures. This problem has been addressed by a minor change in the algorithm (see the companion publication, Blood Press Monit, 6, 161-165, 2001). CONCLUSION: The BPM-100(Beta) is an accurate blood pressure monitor for the office setting, meeting all requirements of the Association for the Advancement of Medical Instrumentation and achieving an 'A' grade according to the British Hypertension Society protocol.
Key concepts: Medicine, Sphygmomanometer, Blood pressure, Heart rate, Sitting, Cardiology, Medical instrumentation, Internal medicine