1993Analytical LettersRequires access

Improved High Performance Liquid Chromatographic Determination of Chlorzoxazone and its Hydroxy Metabolite in Human Serum and Urine

Hailang L. Zhang, James T. Stewart

Open publisher page 9 citations

Abstract

Methods for the high performance liquid chromatographic determinations of chlorzoxazone and its 6-hydroxy metabolite in human serum and urine are presented. The separation of the analytes is achieved within 15 min on an octadecylsilane column with a mobile phase of 30:70 v/v acetonitrile - aqueous 0.05 M sodium dihydrogen phosphate at a pH 4.4 flow rate of 1 mL/min. The serum procedure utilized an octadecylsilane solid-phase extraction clean-up step with injection of the eluent onto the analytical column and detection at 280nm. The urine method uses a buffer dilution of the urine sample and direct injection onto the analytical column with detection at 295nm. Recoveries of chlorzoxazone and metabolite from serum and urine samples are in the 88–103% range. Drug to internal standard peak height ratios are linear in the 0.5 – 50 μg/mL and 5–100 μg/mL ranges for both analytes in serum and urine, respectively. Limits of quantitation for chlorzoxazone are 0.12 μg/mL in serum and 0.62 μg/mL in urine. Limits of quantitation for the 6-hydroxy metabolite are 0.23 μg/mL in serum and 0.029 μg/mL in urine. Accuracy and precision of both methods are in the 0–5% range, respectively.

About this research paper

What this paper is about

Methods for the high performance liquid chromatographic determinations of chlorzoxazone and its 6-hydroxy metabolite in human serum and urine are presented. The separation of the analytes is achieved within 15 min on an octadecylsilane column with a mobile phase of 30:70 v/v acetonitrile - aqueous 0.05 M sodium dihydrogen phosphate at a pH 4.4 flow rate of 1 mL/min. The serum procedure utilized an octadecylsilane solid-phase extraction clean-up step with injection of the eluent onto the analytical column and detection at 280nm. The urine method uses a buffer dilution of the urine sample and direct injection onto the analytical column with detection at 295nm. Recoveries of chlorzoxazone and metabolite from serum and urine samples are in the 88–103% range. Drug to internal standard peak height ratios are linear in the 0.5 – 50 μg/mL and 5–100 μg/mL ranges for both analytes in serum and urine, respectively. Limits of quantitation for chlorzoxazone are 0.12 μg/mL in serum and 0.62 μg/mL in urine. Limits of quantitation for the 6-hydroxy metabolite are 0.23 μg/mL in serum and 0.029 μg/mL in urine. Accuracy and precision of both methods are in the 0–5% range, respectively.

Why it matters

OpenAlex reports 9 citations for this work. Citation counts describe recorded attention and do not establish research quality.

Key contribution

A contribution statement is not available in the OpenAlex record.

Method / approach

Method details are not available in the OpenAlex metadata.

Main findings

Findings are not separately available in the OpenAlex metadata.

Limitations

Limitations are not available in the OpenAlex metadata.

Applications

Application details are not available in the OpenAlex metadata.

Available abstract

Methods for the high performance liquid chromatographic determinations of chlorzoxazone and its 6-hydroxy metabolite in human serum and urine are presented. The separation of the analytes is achieved within 15 min on an octadecylsilane column with a mobile phase of 30:70 v/v acetonitrile - aqueous 0.05 M sodium dihydrogen phosphate at a pH 4.4 flow rate of 1 mL/min. The serum procedure utilized an octadecylsilane solid-phase extraction clean-up step with injection of the eluent onto the analytical column and detection at 280nm. The urine method uses a buffer dilution of the urine sample and direct injection onto the analytical column with detection at 295nm. Recoveries of chlorzoxazone and metabolite from serum and urine samples are in the 88–103% range. Drug to internal standard peak height ratios are linear in the 0.5 – 50 μg/mL and 5–100 μg/mL ranges for both analytes in serum and urine, respectively. Limits of quantitation for chlorzoxazone are 0.12 μg/mL in serum and 0.62 μg/mL in urine. Limits of quantitation for the 6-hydroxy metabolite are 0.23 μg/mL in serum and 0.029 μg/mL in urine. Accuracy and precision of both methods are in the 0–5% range, respectively.

Key concepts: Chemistry, Chromatography, Chlorzoxazone, Metabolite, Urine, Detection limit, Analyte, High-performance liquid chromatography

Related papers

Back to paper searchBrowse research topicsOriginal source
Improved High Performance Liquid Chromatographic Determination of Chlorzoxazone and its Hydroxy Metabolite in Human Serum and Urine — Research Paper | ScholarLens