RELIABILITY AND VALIDITY OF VENOUS BLOOD LACTATE MEASUREMENT WITH ACCUSPORT PORTABLE LACTATE ANALYZER

R. S. Interdonato, Maurie J. Luetkemeier

Open publisher page 1 citations

Abstract

1785 The goal of this project was to compare venous blood lactate values measured by the Accusport portable lactate analyzer to readings made by the "gold standard" spectrophatometric (Spec) technique. Ten subjects (7 females, 3 males) donated 7-ml of venous blood for analysis. The blood samples were divided and augmented with exogenous lactate to produce six simulated exercise concentrations for each measurement technique. The first level contained no exogenous lactate (0-mmol/L) and the remaining 5 levels were "spiked" with 3, 6, 9, 12, and 15-mmol/L of exogenous lactate from Sigma Chemical Co. (St. Louis, MO). Each sample was measured in duplicate at each level by each technique. Individual paired t-tests were used to determine significant differences between techniques at each level of lactate augmentation. The reliability coefficient for duplicate readings on the Accusport was 0.99. Values from the Spec were significantly higher than Accusport at the +3, +6, and +15-mmol/L levels.(Table)TableThe correlation between venous blood lactate values measured on the Accusport and the Spec was 0.99. The mean difference between readings of the Accusport and the Spec across all levels was −0.7-mm/L. This meets the approval criteria as set forth by the FDA, i.e., ±0.75. Although, the Accusport may slightly underestimate true lactate readings, it is probably an acceptable tool for field test measurements of venous blood lactate.

About this research paper

What this paper is about

1785 The goal of this project was to compare venous blood lactate values measured by the Accusport portable lactate analyzer to readings made by the "gold standard" spectrophatometric (Spec) technique. Ten subjects (7 females, 3 males) donated 7-ml of venous blood for analysis. The blood samples were divided and augmented with exogenous lactate to produce six simulated exercise concentrations for each measurement technique. The first level contained no exogenous lactate (0-mmol/L) and the remaining 5 levels were "spiked" with 3, 6, 9, 12, and 15-mmol/L of exogenous lactate from Sigma Chemical Co. (St. Louis, MO). Each sample was measured in duplicate at each level by each technique. Individual paired t-tests were used to determine significant differences between techniques at each level of lactate augmentation. The reliability coefficient for duplicate readings on the Accusport was 0.99. Values from the Spec were significantly higher than Accusport at the +3, +6, and +15-mmol/L levels.(Table)TableThe correlation between venous blood lactate values measured on the Accusport and the Spec was 0.99. The mean difference between readings of the Accusport and the Spec across all levels was −0.7-mm/L. This meets the approval criteria as set forth by the FDA, i.e., ±0.75. Although, the Accusport may slightly underestimate true lactate readings, it is probably an acceptable tool for field test measurements of venous blood lactate.

Why it matters

OpenAlex reports 1 citations for this work. Citation counts describe recorded attention and do not establish research quality.

Key contribution

A contribution statement is not available in the OpenAlex record.

Method / approach

Method details are not available in the OpenAlex metadata.

Main findings

Findings are not separately available in the OpenAlex metadata.

Limitations

Limitations are not available in the OpenAlex metadata.

Applications

Application details are not available in the OpenAlex metadata.

Available abstract

1785 The goal of this project was to compare venous blood lactate values measured by the Accusport portable lactate analyzer to readings made by the "gold standard" spectrophatometric (Spec) technique. Ten subjects (7 females, 3 males) donated 7-ml of venous blood for analysis. The blood samples were divided and augmented with exogenous lactate to produce six simulated exercise concentrations for each measurement technique. The first level contained no exogenous lactate (0-mmol/L) and the remaining 5 levels were "spiked" with 3, 6, 9, 12, and 15-mmol/L of exogenous lactate from Sigma Chemical Co. (St. Louis, MO). Each sample was measured in duplicate at each level by each technique. Individual paired t-tests were used to determine significant differences between techniques at each level of lactate augmentation. The reliability coefficient for duplicate readings on the Accusport was 0.99. Values from the Spec were significantly higher than Accusport at the +3, +6, and +15-mmol/L levels.(Table)TableThe correlation between venous blood lactate values measured on the Accusport and the Spec was 0.99. The mean difference between readings of the Accusport and the Spec across all levels was −0.7-mm/L. This meets the approval criteria as set forth by the FDA, i.e., ±0.75. Although, the Accusport may slightly underestimate true lactate readings, it is probably an acceptable tool for field test measurements of venous blood lactate.

Key concepts: Venous blood, Blood lactate, Coefficient of variation, Medicine, Limits of agreement, Spectrum analyzer, Chemistry, Internal medicine

Related papers

Back to paper searchBrowse research topicsOriginal source
RELIABILITY AND VALIDITY OF VENOUS BLOOD LACTATE MEASUREMENT WITH ACCUSPORT PORTABLE LACTATE ANALYZER — Research Paper | ScholarLens