Reduced airway resistance and work of breathing during mechanical ventilation with an ultra-thin, two-stage polyurethane endotracheal tube (the Kolobow tube)

Carlos A. Velarde, Billie Lou Short, Oswaldo Rivera, Winslow R. Seale, Robert Howard, Theodor Kolobow

Open publisher page 5 citations

Abstract

OBJECTIVES: To compare dynamic pulmonary function studies using the ultrathin walled Kolobow endotracheal tube, with conventional endotracheal tubes of similar external diameter on rabbits during mechanical ventilation. To test the hypothesis that the increased internal diameter of the Kolobow tube will result in decreased airway resistance and work of breathing. DESIGN: Controlled animal study. SETTING: Institutional animal research facility. SUBJECTS: Adult female Dutch Belted rabbits (n = 6), weighing 1.4 to 1.6 kg. INTERVENTIONS: The animals were initially intubated with a conventional endotracheal tube (2.5-mm internal diameter; 3.6-mm outer diameter); they were paralyzed and placed on a mechanical ventilator. Ventilatory settings were adjusted to obtain standard arterial blood gases: pH of 7.35 to 7.45; PaCO2 of 35 to 40 torr (4.7 to 5.3 kPa), and PaO2 of 90 to 100 torr (12.0 to 13.3 kPa). After the stabilization period, pulmonary function tests (PFTs) were measured (period 1), the conventional endotracheal tube was replaced with a Kolobow tube, and PFTs were measured again and recorded (period 2). While continuously monitoring tidal volume, the peak inspiratory pressure was decreased to match the tidal volume measured during ventilation with the conventional endotracheal tube. Once the desired tidal volume was reached, PFTs were recorded (period 3). Flows were unchanged during the experiment and the length of the endotracheal tubes was the same for both the conventional and the Kolobow tube. MEASUREMENTS AND MAIN RESULTS: Mean values of the airway resistance and work of breathing from periods 1 and 3 were compared using the Student's t-test. There was a 59% decrease in total airway resistance (p = .001) and 45% decrease in the work of breathing (p = .0006). CONCLUSIONS: The use of the ultrathin walled Kolobow endotracheal tube resulted in significant decreases in airway resistance and work of breathing, which has the potential for improving the ventilatory mechanics in very small premature newborns.

About this research paper

What this paper is about

OBJECTIVES: To compare dynamic pulmonary function studies using the ultrathin walled Kolobow endotracheal tube, with conventional endotracheal tubes of similar external diameter on rabbits during mechanical ventilation. To test the hypothesis that the increased internal diameter of the Kolobow tube will result in decreased airway resistance and work of breathing. DESIGN: Controlled animal study. SETTING: Institutional animal research facility. SUBJECTS: Adult female Dutch Belted rabbits (n = 6), weighing 1.4 to 1.6 kg. INTERVENTIONS: The animals were initially intubated with a conventional endotracheal tube (2.5-mm internal diameter; 3.6-mm outer diameter); they were paralyzed and placed on a mechanical ventilator. Ventilatory settings were adjusted to obtain standard arterial blood gases: pH of 7.35 to 7.45; PaCO2 of 35 to 40 torr (4.7 to 5.3 kPa), and PaO2 of 90 to 100 torr (12.0 to 13.3 kPa). After the stabilization period, pulmonary function tests (PFTs) were measured (period 1), the conventional endotracheal tube was replaced with a Kolobow tube, and PFTs were measured again and recorded (period 2). While continuously monitoring tidal volume, the peak inspiratory pressure was decreased to match the tidal volume measured during ventilation with the conventional endotracheal tube. Once the desired tidal volume was reached, PFTs were recorded (period 3). Flows were unchanged during the experiment and the length of the endotracheal tubes was the same for both the conventional and the Kolobow tube. MEASUREMENTS AND MAIN RESULTS: Mean values of the airway resistance and work of breathing from periods 1 and 3 were compared using the Student's t-test. There was a 59% decrease in total airway resistance (p = .001) and 45% decrease in the work of breathing (p = .0006). CONCLUSIONS: The use of the ultrathin walled Kolobow endotracheal tube resulted in significant decreases in airway resistance and work of breathing, which has the potential for improving the ventilatory mechanics in very small premature newborns.

Why it matters

OpenAlex reports 5 citations for this work. Citation counts describe recorded attention and do not establish research quality.

Key contribution

A contribution statement is not available in the OpenAlex record.

Method / approach

Method details are not available in the OpenAlex metadata.

Main findings

Findings are not separately available in the OpenAlex metadata.

Limitations

Limitations are not available in the OpenAlex metadata.

Applications

Application details are not available in the OpenAlex metadata.

Available abstract

OBJECTIVES: To compare dynamic pulmonary function studies using the ultrathin walled Kolobow endotracheal tube, with conventional endotracheal tubes of similar external diameter on rabbits during mechanical ventilation. To test the hypothesis that the increased internal diameter of the Kolobow tube will result in decreased airway resistance and work of breathing. DESIGN: Controlled animal study. SETTING: Institutional animal research facility. SUBJECTS: Adult female Dutch Belted rabbits (n = 6), weighing 1.4 to 1.6 kg. INTERVENTIONS: The animals were initially intubated with a conventional endotracheal tube (2.5-mm internal diameter; 3.6-mm outer diameter); they were paralyzed and placed on a mechanical ventilator. Ventilatory settings were adjusted to obtain standard arterial blood gases: pH of 7.35 to 7.45; PaCO2 of 35 to 40 torr (4.7 to 5.3 kPa), and PaO2 of 90 to 100 torr (12.0 to 13.3 kPa). After the stabilization period, pulmonary function tests (PFTs) were measured (period 1), the conventional endotracheal tube was replaced with a Kolobow tube, and PFTs were measured again and recorded (period 2). While continuously monitoring tidal volume, the peak inspiratory pressure was decreased to match the tidal volume measured during ventilation with the conventional endotracheal tube. Once the desired tidal volume was reached, PFTs were recorded (period 3). Flows were unchanged during the experiment and the length of the endotracheal tubes was the same for both the conventional and the Kolobow tube. MEASUREMENTS AND MAIN RESULTS: Mean values of the airway resistance and work of breathing from periods 1 and 3 were compared using the Student's t-test. There was a 59% decrease in total airway resistance (p = .001) and 45% decrease in the work of breathing (p = .0006). CONCLUSIONS: The use of the ultrathin walled Kolobow endotracheal tube resulted in significant decreases in airway resistance and work of breathing, which has the potential for improving the ventilatory mechanics in very small premature newborns.

Key concepts: Medicine, Tidal volume, Work of breathing, Endotracheal tube, Ventilation (architecture), Anesthesia, Mechanical ventilation, Tube (container)