2006•Journal of GlaucomaRequires access

Comparison of ICare Tonometer with Goldmann Applanation Tonometer in Glaucoma Patients

Paolo Brusini, Maria Letizia Salvetat, Marco Zeppieri, Claudia Tosoni, Lucia Parisi

Open publisher page 187 citations

Abstract

PURPOSE: To compare the intraocular pressure (IOP) readings taken with the new ICare tonometer and with the Goldmann applanation tonometer (GAT) and to evaluate the influence of central corneal thickness (CCT) on the IOP measurements. PATIENTS AND METHODS: One eye of 178 consecutive patients with primary open-angle glaucoma underwent ultrasonic CCT measurement, followed by IOP evaluation with the GAT and with the ICare tonometer. The deviation of ICare readings from GAT values, corrected according to the Doughty and Zaman formula, was calculated and correlated to CCT by a linear regression model. The agreement between the 2 devices was assessed by use of the Bland-Altman method. RESULTS: The average CCT was 552+/-39 mum. The mean IOP and the mean corrected IOP with GAT were 19.4+/-5.4 mm Hg, and 18.5+/-5.7 mm Hg, respectively. The mean ICare IOP reading was 18.4+/-5.2 mm Hg. The deviations of ICare readings from corrected GAT values were highly correlated with CCT values (r=0.63, P<0.01). Linear regression analysis showed that a CCT change of 10 mum resulted in an ICare reading deviation of 0.7 mm Hg. The Bland-Altman scatter-plot showed a reasonable agreement between the 2 tonometers. CONCLUSIONS: The ICare tonometer can be useful in a routine clinical setting. The IOP readings are quite in accordance with those obtained by GAT. The measurements seemed to be influenced by CCT variations, and thus pachymetry should always be taken into consideration.

About this research paper

What this paper is about

PURPOSE: To compare the intraocular pressure (IOP) readings taken with the new ICare tonometer and with the Goldmann applanation tonometer (GAT) and to evaluate the influence of central corneal thickness (CCT) on the IOP measurements. PATIENTS AND METHODS: One eye of 178 consecutive patients with primary open-angle glaucoma underwent ultrasonic CCT measurement, followed by IOP evaluation with the GAT and with the ICare tonometer. The deviation of ICare readings from GAT values, corrected according to the Doughty and Zaman formula, was calculated and correlated to CCT by a linear regression model. The agreement between the 2 devices was assessed by use of the Bland-Altman method. RESULTS: The average CCT was 552+/-39 mum. The mean IOP and the mean corrected IOP with GAT were 19.4+/-5.4 mm Hg, and 18.5+/-5.7 mm Hg, respectively. The mean ICare IOP reading was 18.4+/-5.2 mm Hg. The deviations of ICare readings from corrected GAT values were highly correlated with CCT values (r=0.63, P<0.01). Linear regression analysis showed that a CCT change of 10 mum resulted in an ICare reading deviation of 0.7 mm Hg. The Bland-Altman scatter-plot showed a reasonable agreement between the 2 tonometers. CONCLUSIONS: The ICare tonometer can be useful in a routine clinical setting. The IOP readings are quite in accordance with those obtained by GAT. The measurements seemed to be influenced by CCT variations, and thus pachymetry should always be taken into consideration.

Why it matters

OpenAlex reports 187 citations for this work. Citation counts describe recorded attention and do not establish research quality.

Key contribution

A contribution statement is not available in the OpenAlex record.

Method / approach

Method details are not available in the OpenAlex metadata.

Main findings

Findings are not separately available in the OpenAlex metadata.

Limitations

Limitations are not available in the OpenAlex metadata.

Applications

Application details are not available in the OpenAlex metadata.

Available abstract

PURPOSE: To compare the intraocular pressure (IOP) readings taken with the new ICare tonometer and with the Goldmann applanation tonometer (GAT) and to evaluate the influence of central corneal thickness (CCT) on the IOP measurements. PATIENTS AND METHODS: One eye of 178 consecutive patients with primary open-angle glaucoma underwent ultrasonic CCT measurement, followed by IOP evaluation with the GAT and with the ICare tonometer. The deviation of ICare readings from GAT values, corrected according to the Doughty and Zaman formula, was calculated and correlated to CCT by a linear regression model. The agreement between the 2 devices was assessed by use of the Bland-Altman method. RESULTS: The average CCT was 552+/-39 mum. The mean IOP and the mean corrected IOP with GAT were 19.4+/-5.4 mm Hg, and 18.5+/-5.7 mm Hg, respectively. The mean ICare IOP reading was 18.4+/-5.2 mm Hg. The deviations of ICare readings from corrected GAT values were highly correlated with CCT values (r=0.63, P<0.01). Linear regression analysis showed that a CCT change of 10 mum resulted in an ICare reading deviation of 0.7 mm Hg. The Bland-Altman scatter-plot showed a reasonable agreement between the 2 tonometers. CONCLUSIONS: The ICare tonometer can be useful in a routine clinical setting. The IOP readings are quite in accordance with those obtained by GAT. The measurements seemed to be influenced by CCT variations, and thus pachymetry should always be taken into consideration.

Key concepts: Goldmann Applanation Tonometer, Intraocular pressure, Ophthalmology, Medicine, Glaucoma, Absolute deviation, Limits of agreement, Open angle glaucoma