Premonochromator for stray light rejection in Raman spectroscopy
Jacob Fellman, Peter Lindblom
Abstract
Jacob Fellman, Peter Lindblom
Abstract
A new apparatus for attenuating the Rayleigh light is described. The main components are two concave gratings mounted in a special nondispersive way. By means of the first grating the entrance slit of the apparatus is focused as an intermediate spectrum upon a sharp-edged movable mirror. This mirror reflects all the spectrum except the Rayleigh light to the second grating, which collects the light to an image of the entrance slit. The result is a sharp cutting of the spectrum at any desired position. The rejection of the Rayleigh line has been measured to less than 10(-5) (optimally 10(-6)). Raman lines can be.measured about 10 cm(-1) from the exciting line. The instrument has been used as a preeliminating unit coupled to a conventional double monochromator. It is expected that a system consisting of the instrument and a good single monochromator can be used for Raman studies.
OpenAlex reports 6 citations for this work. Citation counts describe recorded attention and do not establish research quality.
A contribution statement is not available in the OpenAlex record.
Method details are not available in the OpenAlex metadata.
Findings are not separately available in the OpenAlex metadata.
Limitations are not available in the OpenAlex metadata.
Application details are not available in the OpenAlex metadata.
A new apparatus for attenuating the Rayleigh light is described. The main components are two concave gratings mounted in a special nondispersive way. By means of the first grating the entrance slit of the apparatus is focused as an intermediate spectrum upon a sharp-edged movable mirror. This mirror reflects all the spectrum except the Rayleigh light to the second grating, which collects the light to an image of the entrance slit. The result is a sharp cutting of the spectrum at any desired position. The rejection of the Rayleigh line has been measured to less than 10(-5) (optimally 10(-6)). Raman lines can be.measured about 10 cm(-1) from the exciting line. The instrument has been used as a preeliminating unit coupled to a conventional double monochromator. It is expected that a system consisting of the instrument and a good single monochromator can be used for Raman studies.
Key concepts: Monochromator, Stray light, Optics, Grating, Rayleigh scattering, Raman spectroscopy, Diffraction grating, Line (geometry)