Evaluation of the Olympus AU 400 clinical chemistry analyzer.
A. Bilic, Ivana Alpeza, Ana Stavljenić-Rukavina
Abstract
A. Bilic, Ivana Alpeza, Ana Stavljenić-Rukavina
Abstract
The performance of the Olympus AU 400 clinical chemistry analyzer was evaluated according to the guidelines of the European Committee for Clinical Laboratory Standards. The following analytes were tested: glucose, urea, creatinine, calcium, AST, ALT, CK, LDH, ALP and amylase. The Olympus AU 400 was compared with the Olympus AU 800. Coefficients of correlation showed high correlation between the compared analyzers. Other performances (intra- and inter-assay variation, carry-over and interferences) of the analyzer were satisfactory.
OpenAlex reports 8 citations for this work. Citation counts describe recorded attention and do not establish research quality.
A contribution statement is not available in the OpenAlex record.
Method details are not available in the OpenAlex metadata.
Findings are not separately available in the OpenAlex metadata.
Limitations are not available in the OpenAlex metadata.
Application details are not available in the OpenAlex metadata.
The performance of the Olympus AU 400 clinical chemistry analyzer was evaluated according to the guidelines of the European Committee for Clinical Laboratory Standards. The following analytes were tested: glucose, urea, creatinine, calcium, AST, ALT, CK, LDH, ALP and amylase. The Olympus AU 400 was compared with the Olympus AU 800. Coefficients of correlation showed high correlation between the compared analyzers. Other performances (intra- and inter-assay variation, carry-over and interferences) of the analyzer were satisfactory.
Key concepts: Spectrum analyzer, Chemistry, Analyte, Creatinine, Analytical Chemistry (journal), Chromatography, Biochemistry, Physics