A standard manganin wire pressure gage was examined by comparing it with a recently developed 0.01% CaF/sub 2/ capacitive pressure gage. The effects of the Bridgman water kick'' and intrinsic time constant are clearly shown, and the results punctuate the usual assertion that manganin wire gages are extremely difficult to use approaching the 0.1% level. (auth)
OpenAlex reports 1 citations for this work. Citation counts describe recorded attention and do not establish research quality.
A contribution statement is not available in the OpenAlex record.
Method details are not available in the OpenAlex metadata.
Findings are not separately available in the OpenAlex metadata.
Limitations are not available in the OpenAlex metadata.
Application details are not available in the OpenAlex metadata.
A standard manganin wire pressure gage was examined by comparing it with a recently developed 0.01% CaF/sub 2/ capacitive pressure gage. The effects of the Bridgman water kick'' and intrinsic time constant are clearly shown, and the results punctuate the usual assertion that manganin wire gages are extremely difficult to use approaching the 0.1% level. (auth)