Evaluation of hemoglobin A2 quantitation assay and hemoglobin variant screening by capillary electrophoresis.
Margaret A. Jenkins, J. G. Hendy, Ivor Smith
Abstract
Margaret A. Jenkins, J. G. Hendy, Ivor Smith
Abstract
To critically assess the method of capillary electrophoresis, we examined 97 clinical samples submitted for hemoglobin electrophoresis by both conventional methods (electrophoresis on cellulose acetate and citrate agar for detection of abnormal variants, ion-exchange chromatography for quantification of HbA2, alkali denaturation for quantification of HbF, supravital stains for detection of HbH) and CE. CE was performed using a 72 cm (50 cm to detector) x 50 microns i.d. fused-silica capillary with detection of absorbance at 200 nm. Of the 97 samples examined, 34 contained a hemoglobin variant. Migration time, the difference of the variant peak to that of HbA, was used in an attempt to identify the hemoglobin variant by CE. We found the method to be suitable for the quantification of HbA2, useful as a screening technique for the presence of hemoglobin variants, but unsuitable for the quantification of HbF.
OpenAlex reports 20 citations for this work. Citation counts describe recorded attention and do not establish research quality.
A contribution statement is not available in the OpenAlex record.
Method details are not available in the OpenAlex metadata.
Findings are not separately available in the OpenAlex metadata.
Limitations are not available in the OpenAlex metadata.
Application details are not available in the OpenAlex metadata.
To critically assess the method of capillary electrophoresis, we examined 97 clinical samples submitted for hemoglobin electrophoresis by both conventional methods (electrophoresis on cellulose acetate and citrate agar for detection of abnormal variants, ion-exchange chromatography for quantification of HbA2, alkali denaturation for quantification of HbF, supravital stains for detection of HbH) and CE. CE was performed using a 72 cm (50 cm to detector) x 50 microns i.d. fused-silica capillary with detection of absorbance at 200 nm. Of the 97 samples examined, 34 contained a hemoglobin variant. Migration time, the difference of the variant peak to that of HbA, was used in an attempt to identify the hemoglobin variant by CE. We found the method to be suitable for the quantification of HbA2, useful as a screening technique for the presence of hemoglobin variants, but unsuitable for the quantification of HbF.
Key concepts: Capillary electrophoresis, Chromatography, Hemoglobin variants, Chemistry, Hemoglobin, Hemoglobin A, Hemoglobin A2, Electrophoresis