Quality assurance of lead aprons for radiation protection
Periyasamy Revathy, Shanmukhappa Kaginelli
Abstract
Periyasamy Revathy, Shanmukhappa Kaginelli
Abstract
Radiation protection is essential to minimize the harmful effects of ionizing radiation. Lead equivalent apron is one of the common protective devices to reduce the occupational risk of radiation exposure. The aim of the study is to analyse the percentage attenuation of 0.5 mm lead equivalent apron. Twenty aprons were used for calculating the percentage of attenuation. The computed tomography (CT) phantom was placed on the radiography and CT table at 1 m from the X-ray tube. The readings were taken from the phantom using a detector at 1-m distance with and without lead equivalent aprons. The readings were tabulated and analysed. The mean percentage attenuation of 0.5 mm lead equivalent apron is 95.66 ± 0.61 in radiography (100 kVp), 95.64 ± 0.57 in CT (130 kVp), respectively. The percentage attenuation of lead equivalent aprons is permissible comparing with the literature.
OpenAlex reports 2 citations for this work. Citation counts describe recorded attention and do not establish research quality.
A contribution statement is not available in the OpenAlex record.
Method details are not available in the OpenAlex metadata.
Findings are not separately available in the OpenAlex metadata.
Limitations are not available in the OpenAlex metadata.
Application details are not available in the OpenAlex metadata.
Radiation protection is essential to minimize the harmful effects of ionizing radiation. Lead equivalent apron is one of the common protective devices to reduce the occupational risk of radiation exposure. The aim of the study is to analyse the percentage attenuation of 0.5 mm lead equivalent apron. Twenty aprons were used for calculating the percentage of attenuation. The computed tomography (CT) phantom was placed on the radiography and CT table at 1 m from the X-ray tube. The readings were taken from the phantom using a detector at 1-m distance with and without lead equivalent aprons. The readings were tabulated and analysed. The mean percentage attenuation of 0.5 mm lead equivalent apron is 95.66 ± 0.61 in radiography (100 kVp), 95.64 ± 0.57 in CT (130 kVp), respectively. The percentage attenuation of lead equivalent aprons is permissible comparing with the literature.
Key concepts: Lead apron, Imaging phantom, Equivalent dose, Radiation protection, Nuclear medicine, Attenuation, Quality assurance, Radiation exposure