Improved peak-to-peak method for cavity length measurement of a Fabry-Perot etalon using a mode-locked femtosecond laser

Dong Wook Shin, Hiraku Matsukuma, Ryo Sato, Eberhard Manske, Wei Gao

Open full text 10 citations

Abstract

Differing from the conventional peak-to-peak method using two neighboring spectral peaks in the frequency-domain fringe spectrum of the spectral response of a Fabry-Perot etalon to a femtosecond laser, which contains N spectral peaks equally spaced with a spacing of the etalon free spectral range (FSR), the proposed method employs a pair of spectral peaks with a spacing of an integer multiple k (k ≫ 1) of FSR for measurement of the etalon cavity length d with a reduced measurement error. Under the constrain of the total N spectral peaks obtainable in the finite spectral range of the femtosecond laser, the optimized k is identified to be N∕2 in consideration of an averaging operation using N - k samples of d to achieve the minimum measurement error. The feasibility of the proposed method is demonstrated by experimental results with an uncertainty analysis based on "Guides to the Expression of Uncertainty in Measurement".

About this research paper

What this paper is about

Differing from the conventional peak-to-peak method using two neighboring spectral peaks in the frequency-domain fringe spectrum of the spectral response of a Fabry-Perot etalon to a femtosecond laser, which contains N spectral peaks equally spaced with a spacing of the etalon free spectral range (FSR), the proposed method employs a pair of spectral peaks with a spacing of an integer multiple k (k ≫ 1) of FSR for measurement of the etalon cavity length d with a reduced measurement error. Under the constrain of the total N spectral peaks obtainable in the finite spectral range of the femtosecond laser, the optimized k is identified to be N∕2 in consideration of an averaging operation using N - k samples of d to achieve the minimum measurement error. The feasibility of the proposed method is demonstrated by experimental results with an uncertainty analysis based on "Guides to the Expression of Uncertainty in Measurement".

Why it matters

OpenAlex reports 10 citations for this work. Citation counts describe recorded attention and do not establish research quality.

Key contribution

A contribution statement is not available in the OpenAlex record.

Method / approach

Method details are not available in the OpenAlex metadata.

Main findings

Findings are not separately available in the OpenAlex metadata.

Limitations

Limitations are not available in the OpenAlex metadata.

Applications

Application details are not available in the OpenAlex metadata.

Available abstract

Differing from the conventional peak-to-peak method using two neighboring spectral peaks in the frequency-domain fringe spectrum of the spectral response of a Fabry-Perot etalon to a femtosecond laser, which contains N spectral peaks equally spaced with a spacing of the etalon free spectral range (FSR), the proposed method employs a pair of spectral peaks with a spacing of an integer multiple k (k ≫ 1) of FSR for measurement of the etalon cavity length d with a reduced measurement error. Under the constrain of the total N spectral peaks obtainable in the finite spectral range of the femtosecond laser, the optimized k is identified to be N∕2 in consideration of an averaging operation using N - k samples of d to achieve the minimum measurement error. The feasibility of the proposed method is demonstrated by experimental results with an uncertainty analysis based on "Guides to the Expression of Uncertainty in Measurement".

Key concepts: Fabry–Pérot interferometer, Optics, Free spectral range, Laser, Femtosecond, Materials science, Range (aeronautics), Spectral line