Intracranial pressure measurement from the anterior fontanelle utilizing a pneumoelectronic switch

L. Bunegin, Maurice S. Albin, R. Rauschhuber, Arthur E. Marlin

Open publisher page 0 citations

Abstract

A newly developed infant cranial model shows that accurate, reproducible, and noninvasive measurements of intracranial pressure (ICP) can be made from the anterior fontanelle when fontanelle pressure is referenced from the bony margins adjacent to the fontanelle opening. Also, this model provides insight into the elastic properties of the fontanelle membrane and the pressure/volume relationships governing ICP transmission through the fontanelle window. An anterior fontanelle pressure monitor design based on data from the infant cranial model was used to monitor anterior fontanelle pressure in three infants with elevated ICP and previously inserted ventricular catheters. Measured anterior fontanelle pressure was highly correlated to ICP (r = 0.962) with high reproducibility after blind application and reapplication. Base line adjustment and in situ recalibration were easily achieved, with the monitor showing no sensitivity to patient movement and excellent frequency response. (Neurosurgery 20:726-731, 1987)

About this research paper

What this paper is about

A newly developed infant cranial model shows that accurate, reproducible, and noninvasive measurements of intracranial pressure (ICP) can be made from the anterior fontanelle when fontanelle pressure is referenced from the bony margins adjacent to the fontanelle opening. Also, this model provides insight into the elastic properties of the fontanelle membrane and the pressure/volume relationships governing ICP transmission through the fontanelle window. An anterior fontanelle pressure monitor design based on data from the infant cranial model was used to monitor anterior fontanelle pressure in three infants with elevated ICP and previously inserted ventricular catheters. Measured anterior fontanelle pressure was highly correlated to ICP (r = 0.962) with high reproducibility after blind application and reapplication. Base line adjustment and in situ recalibration were easily achieved, with the monitor showing no sensitivity to patient movement and excellent frequency response. (Neurosurgery 20:726-731, 1987)

Why it matters

A significance statement is not available in the OpenAlex record.

Key contribution

A contribution statement is not available in the OpenAlex record.

Method / approach

Method details are not available in the OpenAlex metadata.

Main findings

Findings are not separately available in the OpenAlex metadata.

Limitations

Limitations are not available in the OpenAlex metadata.

Applications

Application details are not available in the OpenAlex metadata.

Available abstract

A newly developed infant cranial model shows that accurate, reproducible, and noninvasive measurements of intracranial pressure (ICP) can be made from the anterior fontanelle when fontanelle pressure is referenced from the bony margins adjacent to the fontanelle opening. Also, this model provides insight into the elastic properties of the fontanelle membrane and the pressure/volume relationships governing ICP transmission through the fontanelle window. An anterior fontanelle pressure monitor design based on data from the infant cranial model was used to monitor anterior fontanelle pressure in three infants with elevated ICP and previously inserted ventricular catheters. Measured anterior fontanelle pressure was highly correlated to ICP (r = 0.962) with high reproducibility after blind application and reapplication. Base line adjustment and in situ recalibration were easily achieved, with the monitor showing no sensitivity to patient movement and excellent frequency response. (Neurosurgery 20:726-731, 1987)

Key concepts: Anterior fontanelle, Fontanelle, Medicine, Intracranial pressure, Raised intracranial pressure, Surgery, Hydrocephalus

Related papers

Back to paper searchBrowse research topicsOriginal source
Intracranial pressure measurement from the anterior fontanelle utilizing a pneumoelectronic switch — Research Paper | ScholarLens