Predischarge Screening of Neonates with Transcutaneous Bilirubinometer
Thrilok Natarajan, Ramesh Srinivasan, Sumathi Raju, Suvetha Kannapan
Abstract
Thrilok Natarajan, Ramesh Srinivasan, Sumathi Raju, Suvetha Kannapan
Abstract
Objective: To compare the diagnostic accuracy of various transcutaneous bilirubin (TcB) nomograms for predischarge screening. Methods: The paired total serum bilirubin (TSB) and TcB measurements collected in neonates ≥35 weeks and ≥2000 g birth weight were analyzed. BiliCare™ bilirubinometer was used for TcB measurement. We chose the following nomograms for the study: Bhutani nomogram, Maisel's nomogram, Agarwal nomogram, Thakkar nomogram, American Academy of Pediatrics (AAP) nomogram within 3 mg/dl of phototherapy cutoff, AAP nomogram >70% of phototherapy cutoff and if TcB value is above 13 mg/dl. The diagnostic accuracy of these nomograms for TcB was compared with TSB plotted in the Bhutani nomogram. Results: TcB showed a positive correlation with TSB (Pearson correlation coefficient = 0.783). Bhutani nomogram, Maisel's nomogram and AAP (using within 3 mg/dL cutoff) nomogram showed good sensitivity and low false-negative rate while avoiding blood draws in most neonates. Conclusion: Bhutani nomogram, Maisel's nomogram, and AAP (using within 3 mg/dL of phototherapy cutoff) nomograms have comparable diagnostic accuracy for predischarge bilirubin screening in neonates.
A significance statement is not available in the OpenAlex record.
A contribution statement is not available in the OpenAlex record.
Method details are not available in the OpenAlex metadata.
Findings are not separately available in the OpenAlex metadata.
Limitations are not available in the OpenAlex metadata.
Application details are not available in the OpenAlex metadata.
Objective: To compare the diagnostic accuracy of various transcutaneous bilirubin (TcB) nomograms for predischarge screening. Methods: The paired total serum bilirubin (TSB) and TcB measurements collected in neonates ≥35 weeks and ≥2000 g birth weight were analyzed. BiliCare™ bilirubinometer was used for TcB measurement. We chose the following nomograms for the study: Bhutani nomogram, Maisel's nomogram, Agarwal nomogram, Thakkar nomogram, American Academy of Pediatrics (AAP) nomogram within 3 mg/dl of phototherapy cutoff, AAP nomogram >70% of phototherapy cutoff and if TcB value is above 13 mg/dl. The diagnostic accuracy of these nomograms for TcB was compared with TSB plotted in the Bhutani nomogram. Results: TcB showed a positive correlation with TSB (Pearson correlation coefficient = 0.783). Bhutani nomogram, Maisel's nomogram and AAP (using within 3 mg/dL cutoff) nomogram showed good sensitivity and low false-negative rate while avoiding blood draws in most neonates. Conclusion: Bhutani nomogram, Maisel's nomogram, and AAP (using within 3 mg/dL of phototherapy cutoff) nomograms have comparable diagnostic accuracy for predischarge bilirubin screening in neonates.
Key concepts: Nomogram, Medicine, Cutoff, Bilirubin, Internal medicine, Physics, Quantum mechanics