Performance Evaluation of the Quantitative Point-of-Care LumiraDx d-Dimer Test

Jayne E. Ellis, Thomas W. Johnston, David Wesley Craig, Anita Scribner, William Escudero Simon, Judith Lee Kirstein

Open full text 9 citations

Abstract

INTRODUCTION: Fibrin degradation product D-dimer can be a valuable indicator for venous thromboembolism (VTE). The use of D-dimer testing in primary care settings can be limited by restricted access to laboratory services. This performance evaluation compares a quantitative, point-of-care (POC) D-dimer assay (LumiraDx D-Dimer Test) with a reference laboratory-based D-dimer assay. METHODS: Plasma samples from patients presenting to secondary care in the UK, USA, and Germany were analyzed centrally using the LumiraDx D-Dimer Test and the reference test (bioMérieux VIDAS D-Dimer Exclusion II immunoassay). Method comparison used Passing-Bablok regression analysis with pre-specified equivalence criteria of r ≥ 0.9 and slope of 0.9-1.1. The NOVEL-3 study (NCT04375982) compared equivalency of fingerstick, venous blood (VB), and plasma samples from the same patient, tested at US primary care clinics next to the patient using the POC LumiraDx D-Dimer device. Measurements obtained from fingerstick and VB samples were compared with results from plasma samples, using Deming regression. The healthy reference range was determined using plasma samples of healthy volunteers, collected by commercial suppliers in Germany and the USA, which were analyzed centrally using the LumiraDx D-Dimer Test and the reference test. RESULTS: The LumiraDx D-Dimer Test demonstrated agreement with the bioMérieux VIDAS D-Dimer Exclusion II immunoassay for plasma samples (r = 0.923, slope of 1.016, n = 1767). There was good agreement between fingerstick/VB samples and plasma samples (r = 0.980-0.986, n = 93) measured using the LumiraDx D-Dimer Test. Overall error rates were 1.8%. The healthy reference range 90% percentile for D-dimer was calculated as 533 µg/l fibrinogen equivalent units (FEU). CONCLUSIONS: The quantitative LumiraDx D-Dimer Test is easy to use and can accurately measure D-dimer levels in a range of blood sample types, including fingerstick samples, which could improve assessment of VTE cases in community and hospital near-patient settings.

Open-access reader

About this research paper

What this paper is about

INTRODUCTION: Fibrin degradation product D-dimer can be a valuable indicator for venous thromboembolism (VTE). The use of D-dimer testing in primary care settings can be limited by restricted access to laboratory services. This performance evaluation compares a quantitative, point-of-care (POC) D-dimer assay (LumiraDx D-Dimer Test) with a reference laboratory-based D-dimer assay. METHODS: Plasma samples from patients presenting to secondary care in the UK, USA, and Germany were analyzed centrally using the LumiraDx D-Dimer Test and the reference test (bioMérieux VIDAS D-Dimer Exclusion II immunoassay). Method comparison used Passing-Bablok regression analysis with pre-specified equivalence criteria of r ≥ 0.9 and slope of 0.9-1.1. The NOVEL-3 study (NCT04375982) compared equivalency of fingerstick, venous blood (VB), and plasma samples from the same patient, tested at US primary care clinics next to the patient using the POC LumiraDx D-Dimer device. Measurements obtained from fingerstick and VB samples were compared with results from plasma samples, using Deming regression. The healthy reference range was determined using plasma samples of healthy volunteers, collected by commercial suppliers in Germany and the USA, which were analyzed centrally using the LumiraDx D-Dimer Test and the reference test. RESULTS: The LumiraDx D-Dimer Test demonstrated agreement with the bioMérieux VIDAS D-Dimer Exclusion II immunoassay for plasma samples (r = 0.923, slope of 1.016, n = 1767). There was good agreement between fingerstick/VB samples and plasma samples (r = 0.980-0.986, n = 93) measured using the LumiraDx D-Dimer Test. Overall error rates were 1.8%. The healthy reference range 90% percentile for D-dimer was calculated as 533 µg/l fibrinogen equivalent units (FEU). CONCLUSIONS: The quantitative LumiraDx D-Dimer Test is easy to use and can accurately measure D-dimer levels in a range of blood sample types, including fingerstick samples, which could improve assessment of VTE cases in community and hospital near-patient settings.

Why it matters

OpenAlex reports 9 citations for this work. Citation counts describe recorded attention and do not establish research quality.

Key contribution

A contribution statement is not available in the OpenAlex record.

Method / approach

Method details are not available in the OpenAlex metadata.

Main findings

Findings are not separately available in the OpenAlex metadata.

Limitations

Limitations are not available in the OpenAlex metadata.

Applications

Application details are not available in the OpenAlex metadata.

Available abstract

INTRODUCTION: Fibrin degradation product D-dimer can be a valuable indicator for venous thromboembolism (VTE). The use of D-dimer testing in primary care settings can be limited by restricted access to laboratory services. This performance evaluation compares a quantitative, point-of-care (POC) D-dimer assay (LumiraDx D-Dimer Test) with a reference laboratory-based D-dimer assay. METHODS: Plasma samples from patients presenting to secondary care in the UK, USA, and Germany were analyzed centrally using the LumiraDx D-Dimer Test and the reference test (bioMérieux VIDAS D-Dimer Exclusion II immunoassay). Method comparison used Passing-Bablok regression analysis with pre-specified equivalence criteria of r ≥ 0.9 and slope of 0.9-1.1. The NOVEL-3 study (NCT04375982) compared equivalency of fingerstick, venous blood (VB), and plasma samples from the same patient, tested at US primary care clinics next to the patient using the POC LumiraDx D-Dimer device. Measurements obtained from fingerstick and VB samples were compared with results from plasma samples, using Deming regression. The healthy reference range was determined using plasma samples of healthy volunteers, collected by commercial suppliers in Germany and the USA, which were analyzed centrally using the LumiraDx D-Dimer Test and the reference test. RESULTS: The LumiraDx D-Dimer Test demonstrated agreement with the bioMérieux VIDAS D-Dimer Exclusion II immunoassay for plasma samples (r = 0.923, slope of 1.016, n = 1767). There was good agreement between fingerstick/VB samples and plasma samples (r = 0.980-0.986, n = 93) measured using the LumiraDx D-Dimer Test. Overall error rates were 1.8%. The healthy reference range 90% percentile for D-dimer was calculated as 533 µg/l fibrinogen equivalent units (FEU). CONCLUSIONS: The quantitative LumiraDx D-Dimer Test is easy to use and can accurately measure D-dimer levels in a range of blood sample types, including fingerstick samples, which could improve assessment of VTE cases in community and hospital near-patient settings.

Key concepts: D-dimer, Fingerstick, Point of care, Medicine, Point-of-care testing, Reference range, Internal medicine, Immunology

Related papers

Back to paper searchBrowse research topicsOriginal source
Performance Evaluation of the Quantitative Point-of-Care LumiraDx d-Dimer Test — Research Paper | ScholarLens