Determination of camphor in compound menthol nasal drops by ultraviolet spectrophotometry
GU Wei-ying, Heping Zhang, Huang Wei-fen
Abstract
GU Wei-ying, Heping Zhang, Huang Wei-fen
Abstract
Objective To establish a quantitative analysis method for camphor in compound menthol nasal drops.Methods The camphor was determined directly at (292±1)nm by ultraviolet spectrophotometry.Results:The calibration curve was linear in the range from 1.62~3.24mg/mL with r=0.9992,the average recovery and relative standard deviation were 100.06% and 1.29% respectively. Conclusion The method is simple,rapid and accurate.
A significance statement is not available in the OpenAlex record.
A contribution statement is not available in the OpenAlex record.
Method details are not available in the OpenAlex metadata.
Findings are not separately available in the OpenAlex metadata.
Limitations are not available in the OpenAlex metadata.
Application details are not available in the OpenAlex metadata.
Objective To establish a quantitative analysis method for camphor in compound menthol nasal drops.Methods The camphor was determined directly at (292±1)nm by ultraviolet spectrophotometry.Results:The calibration curve was linear in the range from 1.62~3.24mg/mL with r=0.9992,the average recovery and relative standard deviation were 100.06% and 1.29% respectively. Conclusion The method is simple,rapid and accurate.
Key concepts: Camphor, Menthol, Spectrophotometry, Chromatography, Relative standard deviation, Calibration curve, Chemistry, Ultraviolet