Investigation of absolute dose calibration accuracy for TomoTherapy using real water
Tingtian Pang, Bo Yang, Xia Liu, J. Castle, Lang Yu, Nan Liu, Wenbo Li, Tingting Dong, Jie Qiu, Quan Chen
Abstract
Tingtian Pang, Bo Yang, Xia Liu, J. Castle, Lang Yu, Nan Liu, Wenbo Li, Tingting Dong, Jie Qiu, Quan Chen
Abstract
Abstract A systematic bias in TomoTherapy output calibration was reported by the Imaging and Radiation Oncology Core Houston (IROC‐H) after analyzing intensity‐modulated radiation therapy (IMRT) credentialing results from hundreds of TomoTherapy units. Multiple theories were developed to explain this observation. One theory was that the use of a solid water “cheese” phantom instead of real water in the calibration measurement was the culprit. A phantom filled with distilled water was built to investigate whether our TomoTherapy was miscalibrated due to the use of a solid water phantom. A miscalibration of −1.47% was detected on our TomoTherapy unit. It is found that despite following the vendor's updated recommendation on computed tomography (CT) number to density calibration, the cheese phantom was still mapped to a density of 1.028 g/cm3, rather than the 1.01 g/cm3 value reported in literature. When the density of the cheese phantom was modified to 1.01 g/cm3 in the treatment planning system, the measurement also indicated that our TomoTherapy machine was miscalibrated by −1.52%, agreeing with the real water phantom findings. Our single‐institution finding showed that the cheese phantom density assignment can introduce greater than 1% errors in the TomoTherapy absolute dose calibration. It is recommended that the absolute dose calibration for TomoTherapy be performed either in real water or in the cheese phantom with the density in TPS overridden as 1.01 g/cm3.
OpenAlex reports 6 citations for this work. Citation counts describe recorded attention and do not establish research quality.
A contribution statement is not available in the OpenAlex record.
Method details are not available in the OpenAlex metadata.
Findings are not separately available in the OpenAlex metadata.
Limitations are not available in the OpenAlex metadata.
Application details are not available in the OpenAlex metadata.
Abstract A systematic bias in TomoTherapy output calibration was reported by the Imaging and Radiation Oncology Core Houston (IROC‐H) after analyzing intensity‐modulated radiation therapy (IMRT) credentialing results from hundreds of TomoTherapy units. Multiple theories were developed to explain this observation. One theory was that the use of a solid water “cheese” phantom instead of real water in the calibration measurement was the culprit. A phantom filled with distilled water was built to investigate whether our TomoTherapy was miscalibrated due to the use of a solid water phantom. A miscalibration of −1.47% was detected on our TomoTherapy unit. It is found that despite following the vendor's updated recommendation on computed tomography (CT) number to density calibration, the cheese phantom was still mapped to a density of 1.028 g/cm3, rather than the 1.01 g/cm3 value reported in literature. When the density of the cheese phantom was modified to 1.01 g/cm3 in the treatment planning system, the measurement also indicated that our TomoTherapy machine was miscalibrated by −1.52%, agreeing with the real water phantom findings. Our single‐institution finding showed that the cheese phantom density assignment can introduce greater than 1% errors in the TomoTherapy absolute dose calibration. It is recommended that the absolute dose calibration for TomoTherapy be performed either in real water or in the cheese phantom with the density in TPS overridden as 1.01 g/cm3.
Key concepts: Tomotherapy, Imaging phantom, Calibration, Nuclear medicine, Medicine, Dosimetry, Medical physics, Biomedical engineering