2021•BMC Medical ImagingOpen access

Impact of time of flight and point spread function on quantitative parameters of lung lesions in 18F-FDG PET/CT

Kemin Huang, Yanlin Feng, Weitang Liang, Lin Li

Open full text 0 citations

Abstract

Abstract Background Image reconstruction algorithm is one of the important factors affecting the quantitative parameters of PET/CT. The purpose of this study was to investigate the effects of time of flight (TOF) and point spread function (PSF) on quantitative parameters of lung lesions in 18F-FDG PET/CT. Methods This retrospective study evaluated 60 lung lesions in 39 patients who had undergone 18F-fluoro-deoxy-glucose (FDG) PET/CT. All lesions larger than 10 mm in diameter were included in the study. The PET data were reconstructed with a baseline ordered-subsets expectation–maximization (OSEM) algorithm, OSEM + PSF, OSEM + TOF and OSEM + TOF + PSF respectively. The differences of maximum standard uptake value (SUVmax), mean standard uptake value (SUVmean), metabolic tumor volume (MTV), total lesion glycolysis (TLG)and signal to noise ratio (SNR)were compared among different reconstruction algorithms. Results Compared with OSEM reconstruction, using OSEM + TOF + PSF increased SUVmean and SUVmax by 23.73% and 22.71% respectively, and SNR increased by 70.18%, MTV decreased by 23.84% (p < 0.01). The percentage difference was significantly higher in smaller lesions (diameter 10–22 mm) than in larger lesions (diameter 23–44 mm), and significantly higher in low contrast lesions (SNR ≤ 15.31) than in high contrast lesions (SNR > 15.31). The difference of TLG among various reconstruction algorithms is relatively small, the highest value is − 6.48% of OSEM + TOF + PSF, and the lowest value is 0.81% of OSEM + TOF. Conclusion TOF and PSF significantly affected the quantitative parameters of lung lesions in 18F-FDG PET/CT. OSEM + TOF + PSF can significantly increased SUVmax, SUVmean and SNR, and significantly reduce MTV, especially in small lesions and low contrast lesions. TLG can be relatively stable in different reconstruction algorithms.

Open-access reader

About this research paper

What this paper is about

Abstract Background Image reconstruction algorithm is one of the important factors affecting the quantitative parameters of PET/CT. The purpose of this study was to investigate the effects of time of flight (TOF) and point spread function (PSF) on quantitative parameters of lung lesions in 18F-FDG PET/CT. Methods This retrospective study evaluated 60 lung lesions in 39 patients who had undergone 18F-fluoro-deoxy-glucose (FDG) PET/CT. All lesions larger than 10 mm in diameter were included in the study. The PET data were reconstructed with a baseline ordered-subsets expectation–maximization (OSEM) algorithm, OSEM + PSF, OSEM + TOF and OSEM + TOF + PSF respectively. The differences of maximum standard uptake value (SUVmax), mean standard uptake value (SUVmean), metabolic tumor volume (MTV), total lesion glycolysis (TLG)and signal to noise ratio (SNR)were compared among different reconstruction algorithms. Results Compared with OSEM reconstruction, using OSEM + TOF + PSF increased SUVmean and SUVmax by 23.73% and 22.71% respectively, and SNR increased by 70.18%, MTV decreased by 23.84% (p < 0.01). The percentage difference was significantly higher in smaller lesions (diameter 10–22 mm) than in larger lesions (diameter 23–44 mm), and significantly higher in low contrast lesions (SNR ≤ 15.31) than in high contrast lesions (SNR > 15.31). The difference of TLG among various reconstruction algorithms is relatively small, the highest value is − 6.48% of OSEM + TOF + PSF, and the lowest value is 0.81% of OSEM + TOF. Conclusion TOF and PSF significantly affected the quantitative parameters of lung lesions in 18F-FDG PET/CT. OSEM + TOF + PSF can significantly increased SUVmax, SUVmean and SNR, and significantly reduce MTV, especially in small lesions and low contrast lesions. TLG can be relatively stable in different reconstruction algorithms.

Why it matters

A significance statement is not available in the OpenAlex record.

Key contribution

A contribution statement is not available in the OpenAlex record.

Method / approach

Method details are not available in the OpenAlex metadata.

Main findings

Findings are not separately available in the OpenAlex metadata.

Limitations

Limitations are not available in the OpenAlex metadata.

Applications

Application details are not available in the OpenAlex metadata.

Available abstract

Abstract Background Image reconstruction algorithm is one of the important factors affecting the quantitative parameters of PET/CT. The purpose of this study was to investigate the effects of time of flight (TOF) and point spread function (PSF) on quantitative parameters of lung lesions in 18F-FDG PET/CT. Methods This retrospective study evaluated 60 lung lesions in 39 patients who had undergone 18F-fluoro-deoxy-glucose (FDG) PET/CT. All lesions larger than 10 mm in diameter were included in the study. The PET data were reconstructed with a baseline ordered-subsets expectation–maximization (OSEM) algorithm, OSEM + PSF, OSEM + TOF and OSEM + TOF + PSF respectively. The differences of maximum standard uptake value (SUVmax), mean standard uptake value (SUVmean), metabolic tumor volume (MTV), total lesion glycolysis (TLG)and signal to noise ratio (SNR)were compared among different reconstruction algorithms. Results Compared with OSEM reconstruction, using OSEM + TOF + PSF increased SUVmean and SUVmax by 23.73% and 22.71% respectively, and SNR increased by 70.18%, MTV decreased by 23.84% (p < 0.01). The percentage difference was significantly higher in smaller lesions (diameter 10–22 mm) than in larger lesions (diameter 23–44 mm), and significantly higher in low contrast lesions (SNR ≤ 15.31) than in high contrast lesions (SNR > 15.31). The difference of TLG among various reconstruction algorithms is relatively small, the highest value is − 6.48% of OSEM + TOF + PSF, and the lowest value is 0.81% of OSEM + TOF. Conclusion TOF and PSF significantly affected the quantitative parameters of lung lesions in 18F-FDG PET/CT. OSEM + TOF + PSF can significantly increased SUVmax, SUVmean and SNR, and significantly reduce MTV, especially in small lesions and low contrast lesions. TLG can be relatively stable in different reconstruction algorithms.

Key concepts: Nuclear medicine, Point spread function, Positron emission tomography, Algorithm, Lung, Iterative reconstruction, Standardized uptake value, Medicine

Related papers

Back to paper searchBrowse research topicsOriginal source
Impact of time of flight and point spread function on quantitative parameters of lung lesions in 18F-FDG PET/CT — Research Paper | ScholarLens