Performance verification of a new domestic chemiluminescence detection system

Lihua Yang, Lin Wang, Yuji Gong, Wen‐Pu Shi, Xiaohu An, Huifei Zheng

Open full text 5 citations

Abstract

BACKGROUND: Procalcitonin (PCT) is an acute phase response protein, which can be used as an indicator for early diagnosis of infection. At present, the main detection methods for PCT are electrochemiluminescence and enzyme-linked immunofluorescence. We aimed to explore the accuracy of PCT determination in a domestic chemiluminescence detection system and its correlation with other systems. METHODS: Clinical specimens were collected, and the precision, linearity, biological reference interval, contamination rate, Clinical reportable scope, and methodological comparison of the determination of PCT in a Chinese chemiluminescence detection system were evaluated and preliminarily verified by referring to Clinical and Laboratory Standards Institute (CLSI) documents or industry standards. RESULTS: The results of precision verification showed that the coefficient of variation (CV) values of the variation coefficient of precision in the samples of low and high values were 2.07% and 0.83% respectively, while the CV values of the total variation coefficient of precision were 3.05% and 1.81% respectively; these findings all met the experimental requirements. The results of linear verification test showed that the linear range was 0.006-96.96 ng/mL, and the linear relationship was well within the detection range (R2 =0.9891). The biological reference interval and the carrying contamination rate were also verified. The clinical reportable range was 0.02-369.585 ng/mL. The results showed that the correlation coefficient between the Mindray CL900I and the Roche E602 was 0.9986, and that between the Mindray CL900I and the Snibe 2000 was 0.983. Meanwhile, when the PCT was higher than 0.1 ng/mL, the correlation coefficient was 100%. CONCLUSIONS: The domestic chemiluminescence detection system has a good performance in the determination of calcitonin, as indicated by the measures of precision, linearity, biological reference interval, carrying contamination rate, and Clinical reportable scope, and can thus be used for clinical specimen detection. The results of methodological comparison showed that the correlation coefficient between the Mindray CL900I and Roche E602 was 0.9986, while the correlation coefficient between the Mindray CL900I and the Snibe 2000 was 0.983. The test results were consistent with the experimental requirements.

Open-access reader

About this research paper

What this paper is about

BACKGROUND: Procalcitonin (PCT) is an acute phase response protein, which can be used as an indicator for early diagnosis of infection. At present, the main detection methods for PCT are electrochemiluminescence and enzyme-linked immunofluorescence. We aimed to explore the accuracy of PCT determination in a domestic chemiluminescence detection system and its correlation with other systems. METHODS: Clinical specimens were collected, and the precision, linearity, biological reference interval, contamination rate, Clinical reportable scope, and methodological comparison of the determination of PCT in a Chinese chemiluminescence detection system were evaluated and preliminarily verified by referring to Clinical and Laboratory Standards Institute (CLSI) documents or industry standards. RESULTS: The results of precision verification showed that the coefficient of variation (CV) values of the variation coefficient of precision in the samples of low and high values were 2.07% and 0.83% respectively, while the CV values of the total variation coefficient of precision were 3.05% and 1.81% respectively; these findings all met the experimental requirements. The results of linear verification test showed that the linear range was 0.006-96.96 ng/mL, and the linear relationship was well within the detection range (R2 =0.9891). The biological reference interval and the carrying contamination rate were also verified. The clinical reportable range was 0.02-369.585 ng/mL. The results showed that the correlation coefficient between the Mindray CL900I and the Roche E602 was 0.9986, and that between the Mindray CL900I and the Snibe 2000 was 0.983. Meanwhile, when the PCT was higher than 0.1 ng/mL, the correlation coefficient was 100%. CONCLUSIONS: The domestic chemiluminescence detection system has a good performance in the determination of calcitonin, as indicated by the measures of precision, linearity, biological reference interval, carrying contamination rate, and Clinical reportable scope, and can thus be used for clinical specimen detection. The results of methodological comparison showed that the correlation coefficient between the Mindray CL900I and Roche E602 was 0.9986, while the correlation coefficient between the Mindray CL900I and the Snibe 2000 was 0.983. The test results were consistent with the experimental requirements.

Why it matters

OpenAlex reports 5 citations for this work. Citation counts describe recorded attention and do not establish research quality.

Key contribution

A contribution statement is not available in the OpenAlex record.

Method / approach

Method details are not available in the OpenAlex metadata.

Main findings

Findings are not separately available in the OpenAlex metadata.

Limitations

Limitations are not available in the OpenAlex metadata.

Applications

Application details are not available in the OpenAlex metadata.

Available abstract

BACKGROUND: Procalcitonin (PCT) is an acute phase response protein, which can be used as an indicator for early diagnosis of infection. At present, the main detection methods for PCT are electrochemiluminescence and enzyme-linked immunofluorescence. We aimed to explore the accuracy of PCT determination in a domestic chemiluminescence detection system and its correlation with other systems. METHODS: Clinical specimens were collected, and the precision, linearity, biological reference interval, contamination rate, Clinical reportable scope, and methodological comparison of the determination of PCT in a Chinese chemiluminescence detection system were evaluated and preliminarily verified by referring to Clinical and Laboratory Standards Institute (CLSI) documents or industry standards. RESULTS: The results of precision verification showed that the coefficient of variation (CV) values of the variation coefficient of precision in the samples of low and high values were 2.07% and 0.83% respectively, while the CV values of the total variation coefficient of precision were 3.05% and 1.81% respectively; these findings all met the experimental requirements. The results of linear verification test showed that the linear range was 0.006-96.96 ng/mL, and the linear relationship was well within the detection range (R2 =0.9891). The biological reference interval and the carrying contamination rate were also verified. The clinical reportable range was 0.02-369.585 ng/mL. The results showed that the correlation coefficient between the Mindray CL900I and the Roche E602 was 0.9986, and that between the Mindray CL900I and the Snibe 2000 was 0.983. Meanwhile, when the PCT was higher than 0.1 ng/mL, the correlation coefficient was 100%. CONCLUSIONS: The domestic chemiluminescence detection system has a good performance in the determination of calcitonin, as indicated by the measures of precision, linearity, biological reference interval, carrying contamination rate, and Clinical reportable scope, and can thus be used for clinical specimen detection. The results of methodological comparison showed that the correlation coefficient between the Mindray CL900I and Roche E602 was 0.9986, while the correlation coefficient between the Mindray CL900I and the Snibe 2000 was 0.983. The test results were consistent with the experimental requirements.